FDA Adverse Event Malfunction Summary report: N

AED PLUS

MDR report key: 4963959 · Received July 30, 2015

Report

Report Number
1220908-2015-01971
Event Type
Malfunction
Date Received
July 30, 2015
Report Date
July 17, 2015
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K011541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A F/U REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 1

THE MALFUNCTION WAS DUPLICATED AND THE MAIN BOARD WAS REPLACED TO RESOLVE THE MALFUNCTION. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
498114 AED PLUS DEFIBRILLATOR MKJ ZOLL MEDICAL CORPORATION AED PLUS NA

Patients

Seq Age Sex Outcome Treatment
1 NA