FDA Adverse Event
Malfunction
Summary report: N
CLASSIC PLUS METZENBAUM SCISSOR
MDR report key: 4963915
·
Received July 30, 2015
Report
- Report Number
- 3007208013-2015-00014
- Event Type
- Malfunction
- Date Received
- July 30, 2015
- Date of Event
- June 10, 2015
- Report Date
- July 9, 2015
- Manufacturer
- SYMMETRY SURGICAL INC.
- Product Code
- LRW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- MEDICAL ASSISTANT
Narratives
Additional Manufacturer Narrative · 1
DEVICE (METZ SCISSORS) WAS EVALUATED BY OUR CONTRACT SUPPLIER, SUPPLIER STATED: THE HARDNESS IS IN THE CENTRAL REGION AND THE MICROSCOPIC EXAMINATION SHOWED A FLAWLESS STRUCTURE. THE BREAK CAN BE CAUSED ONLY BY EXCESSIVE STRESS. ALL MATERIALS AND HARDNESS ARE TO THE CORRECT SPEC.
Description of Event or Problem · 1
THE SCISSOR BROKE INTO TWO PIECES DURING A PROCEDURE, NO PATIENT HARM. 10 MIN DELAY IN SURGERY UNTIL A NEW INSTRUMENT WAS RETRIEVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 498884 | CLASSIC PLUS METZENBAUM SCISSOR | SCISSOR | LRW | SYMMETRY SURGICAL INC. | 36-5016 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |