FDA Adverse Event Malfunction Summary report: N

ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT

MDR report key: 4963728 · Received July 29, 2015

Report

Report Number
1419322-2015-00074
Event Type
Malfunction
Date Received
July 29, 2015
Date of Event
July 1, 2015
Report Date
July 1, 2015
Manufacturer
DENTSPLY PROFESSIONAL
Product Code
EKX
PMA / PMN Number
K031145
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THOUGH NO MEDICAL/SURGICAL INTERVENTION WAS REQUIRED TO PRECLUDE A SERIOUS INJURY IN THIS EVENT, THERE HAVE BEEN PREVIOUSLY REPORTED EVENTS INVOLVING THIS DEVICE THAT RESULTED IN THE NEED FOR MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION. THEREFORE, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE COMPLAINT WAS VERIFIED. MAXIMUM TEMPERATURE EXCEEDED THE TEMPERATURE LIMIT. THE ATTACHMENT EXHIBITED TEMPERATURE INCREASE DURING FREE RUN TESTING 39.1 DEGREE CELSIUS AND LOAD TESTING OF 46.5 CELSIUS. THE EXTERIOR OF THE ATTACHMENT DID NOT SHOW ANY MAJOR WEAR. MICROSCOPIC EVALUATION REVEALED MODERATE GEAR WEAR ON THE HEAD-TAIL AND HEAD ASSEMBLIES AND DETACHMENT AND SUBSEQUENT LODGING OF THE CAP END BEARING IN THE CAP CAVITY. THIS WOULD HAVE LED TO SET INSTABILITY, CONTRIBUTING TO THE OVERHEATING SEEN DURING THE INVESTIGATION. THE INTERIOR OF THE HEAD CAVITY AND GEARS WERE DRY AND DID NOT SHOW SIGNS OF LUBRICATION.

Description of Event or Problem · 1

IN THIS EVENT IT WAS REPORTED THAT AN ESTYLUS 1:5 HANDPIECE OVERHEATED. THERE WAS NO INJURY OR INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
492349 ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT HANDPIECE, DIRECT DRIVE, AC-POWERED EKX DENTSPLY PROFESSIONAL NA

Patients

Seq Age Sex Outcome Treatment
1