FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 4963682 · Received July 28, 2015

Report

Report Number
3006697241-2015-00282
Event Type
Malfunction
Date Received
July 28, 2015
Date of Event
July 2, 2015
Report Date
July 2, 2015
Manufacturer
HILLROM DE MEXICO S DE RL DE CV
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE COMMUNICATION CABLE INOPERABLE. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE COMMUNICATION CABLE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE NURSE CALL WAS INOPERATIVE. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
492099 CAREASSIST BED A/C POWERED ADJUSTABLE HOSP. BED FNL HILLROM DE MEXICO S DE RL DE CV 1170