ARGUS SYSTEM
Report
- Report Number
- 1125873-2015-00013
- Event Type
- Death
- Date Received
- July 31, 2015
- Date of Event
- December 10, 2011
- Report Date
- July 28, 2015
- Manufacturer
- PHILIPS VISICU
- Product Code
- MSX
- PMA / PMN Number
- K001972
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
REVIEW OF DATABASE LOGS WAS USED TO EVALUATE THE SOFTWARE AT THE HEALTH SYSTEM. THE COMPLAINT WAS INVESTIGATED THE DAY IT WAS REPORTED. ONCE THE DATE OF THE EVENT WAS CORRECTED TO BE (B)(6) 2011, THE AVAILABLE SYSTEM LOGS IDENTIFIED THAT THE SOFTWARE BEHAVED ACCORDING TO DESIGN PER THE PREDEFINED HR RULES. THE RULES FOR TRIGGERING AN HR ALERT ARE: 2 CONSECUTIVE 1-MIN HR VALUES > 150 OR 10-MIN HR > 150 OR 2 CONSECUTIVE 1-MIN HR VALUES < 40 OR 10-MIN HR < 40 OR 10-MIN HR VALUE > PT UPPER LIMIT (130) OR 10-MIN HR VALUE < PT LOWER LIMIT (50). THE 5 MINUTE AND 10 MINUTE LOGS INDICATE THAT DEVICE FUNCTIONED AS INTENDED. THE 1 MINUTE LOGS HAD BEEN PURGED OUT OF THE SYSTEM AND USER FACILITY DID NOT PROVIDE THE BACKUP LOGS FOR PHILIPS VISICU TO RESEARCH. THIS REPORT IS BEING FILED AS THE ECARE MANAGER DEVICE WAS USED DURING AN EVENT THAT WAS FOLLOWED BY A PT'S DEATH. ADDITIONAL INFO BECAME KNOWN ON 07/28/2015 DURING A RETROSPECTIVE REVIEW OF COMPLAINTS. PHILIPS VISICU ATTEMPTED TO CONFIRM THE PT'S DEATH WITH USER FACILITY, HOWEVER, THE USER FACILITY WAS UNABLE TO PROVIDE CONFIRMATION. THERE IS NO DATA THAT INDICATES THIS EVENT WAS RELATED TO A DEVICE DEFECT OR MALFUNCTION. ALL AVAILABLE INFO HAS BEEN PLACED ON FILE IN QUALITY ASSURANCE FOR APPROPRIATE TRACKING AND TRENDING.
PHILLIPS VISICU IDENTIFIED A REPORTABLE EVENT DURING A RETROSPECTIVE REVIEW OF COMPLAINTS. THE USER FACILITY REPORTED ON 12/13/2011 THAT THE SMART ALERTS IN ECARE MANAGER DID NOT FIRE ON ONE PT AND THEY SHOULD HAVE. THE USER FACILITY STATED THAT THE HR ALERT SHOULD HAVE FIRED AT 7:06 ON (B)(6) 2011. FURTHER INFO PROVIDED STATED THAT THE PT WAS CODING WHEN THE SMART ALERTS SHOULD HAVE GONE OFF. LATER THE USER FACILITY CONFIRMED THAT THE CORRECT DATE OF THE EVENT WAS (B)(6) 2011. ON 07/28/2015 DURING RETROSPECTIVE REVIEW, IT WAS IDENTIFIED THROUGH PHILIPS VISICU'S DATA LOGS THAT THE PT HAD EXPIRED SOMETIME AFTER THE INCIDENT. WE WERE UNABLE TO CONFIRM THE EXACT TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 499137 | ARGUS SYSTEM | ECAREMANAGER SOFTWARE (V.3.7.1) | MSX | PHILIPS VISICU | 865325 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Death |