PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2015-24807
- Event Type
- Injury
- Date Received
- August 3, 2015
- Date of Event
- July 6, 2015
- Report Date
- July 13, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS
- Reporter Occupation
- PATIENT
Narratives
UNIT PASSED DISPLACEMENT TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/DELIVERY TEST AND EXCESSIVE NO DELIVERY TEST. NO EXCESSIVE NO DELIVERY ALARMS NOTED DURING TESTING. ALL BUTTONS FUNCTIONING PROPERLY. HOWEVER, FOUND CORRODED KEYPAD TRACES. UNIT RECEIVED WITH BROKEN RESERVOIR TUBE LIP, MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, CRACKED DISPLAY WINDOW, MISSING END CAP STICKER AND CRACKED RESERVOIR TUBE.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP'S KEYPAD WAS UNRESPONSIVE. THE CUSTOMER ALSO REPORTED THAT THE DEVICE RECENTLY HAD A NO DELIVERY ALARM. BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS 440 MG/DL, WHICH WAS TREATED WITH A MANUAL INJECTION. THE CUSTOMER REPORTED THAT THE DEVICE WAS WET TO THE TOUCH. CUSTOMER ALSO REPORTED THAT HER BLOOD GLUCOSE LEVEL RECENTLY DROPPED FROM 145 MG/DL TO 59 MG/DL. TROUBLESHOOTING COULD NOT BE PERFORMED AS THE KEYPAD WAS UNRESPONSIVE. CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 504696 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Required Intervention |