FDA Adverse Event Malfunction Summary report: N

PRISMATIC

MDR report key: 4961933 · Received July 31, 2015

Report

Report Number
9710055-2015-00062
Event Type
Malfunction
Date Received
July 31, 2015
Date of Event
July 3, 2015
Report Date
July 3, 2015
Manufacturer
MAQUET S.A.S
Product Code
FTD
PMA / PMN Number
K882613
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A MAQUET (B)(4) VISITED THE FACILITY, AND MET WITH THE CUSTOMER. THEY DETERMINED THAT THIS EVENT IS DUE TO THE AGE OF THE LIGHT (20+ YEARS OLD) AND THE ARMS CONSTANTLY COLLIDING AGAINST ONE ANOTHER FOR MANY YEARS. THE CUSTOMER HAS COVERED AREAS OF THE LIGHT WHERE THE CHIPPING IS AT ITS WORST. THE PRISMATIC SERIES YEARLY PREVENTIVE MAINTENANCE PROGRAM INCLUDES A VERIFICATION OF THE "PAINTING CONDITION, ESPECIALLY IN THE EVENT OF SHOCKS".

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE LIGHT HAS PAINT CHIPPING OFF OF THE SPRING ARM AND MAIN ARM DUE TO IMPACT WITH THE DOWNTUBE AND OTHER MAIN ARMS. THE HOSPITAL HAS BEEN EXPERIENCING SOME PAINT CHIPS FALLING ONTO OPERATION ROOM FLOOR AND INTO THE STERILE FIELD DURING CERTAIN PROCEDURES. NO INJURIES WERE REPORTED. (B)(4). MAQUET IS AWARE OF A TOTAL OF 6 SIMILAR EVENTS INVOLVING 6 SEPARATE DEVICES. THESE 6 EVENTS ARE BEING REPORTED IN THE MDR'S #9810055-2015-00060 THROUGH 9710055-2015-00065.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
498913 PRISMATIC FTD MAQUET S.A.S PRC

Patients

Seq Age Sex Outcome Treatment
1