FDA Adverse Event Malfunction Summary report: N

UNO OPTI-MIST+ NEB 210 CM (10/15) INT

MDR report key: 4961378 · Received July 29, 2015

Report

Report Number
9680866-2015-00071
Event Type
Malfunction
Date Received
July 29, 2015
Report Date
July 3, 2015
Manufacturer
UNOMEDICAL S.A. DE C.V.
Product Code
CAF
PMA / PMN Number
K791536
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PREVIOUS INVESTIGATION HAS BEEN CLOSED. TWO ROOT CAUSES WERE IDENTIFIED. THE JET WAS BEING PRODUCED IN AN OLD INJECTION MOLDING MACHINE THAT WAS NOT CONDITIONED WITH AN ALARM TO DETECT AND PREVENT THE VARIATION (DOUBLE CYCLE) IN THE PROCESS. THIS MACHINE HAS BEEN DISCONTINUED. INADEQUATE INSPECTION PROCESS FOR NEBULIZERS. CORRECTIVE ACTION HAS BEEN IMPLEMENTED TO INCLUDE: VERIFYING ALL ACTIVE MOLDING MACHINES TO CONFIRM ALL HAVE INTEGRATED THE REQUIRED ALARMS. CREATE A TEMPORARY QUALITY ALERT TO INSPECT NEBULIZER COMPONENTS EVERY 2 HOURS AS OUR INSPECTION PROCESS IS UPDATED. CREATION OF A TEST METHOD FOR INSPECTION OF MOLDED COMPONENTS (NEBULIZER SET AND JET). CREATE A TEST METHOD FOR INSPECTION OF MOLDED COMPONENTS, CLARIFY ALL DETAILS OF PROCESS TO DO THE MIXING AND ADD A FORM TO RECORD THE AMOUNT OF RESIN AND MATERIAL USED IN EACH MIXING, DETERMINE WATER FLOW RATES, WATER TEMPERATURE AND PRESSURE, AND INCLUDE PROCESS PARAMETERS TO BE MONITORED. DETERMINED THE EXPECTED SHELF LIFE OF PINS FOR THE ORIFICE OF CUP AND JET, AND ESTABLISH ITS REPLACEMENT IN THE ADEQUATE PERIOD OF TIME. NO ADDITIONAL PATIENT/EVENT DETAILS HAVE BEEN PROVIDED TO DATE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW-UP REPORT WILL BE SUBMITTED. REPORTED TO THE FDA ON DECEMBER 29, 2015. (B)(4).

Additional Manufacturer Narrative · 1

BASED ON THE AVAILABLE INFORMATION, THIS EVENT IS DEEMED A REPORTABLE MALFUNCTION. THERE WAS NO REPORTS OF THE PATIENT HARMED AS A RESULT OF THIS MALFUNCTION. A QUALITY COMPLAINT INVESTIGATION WAS PERFORMED. A USED SAMPLE WAS RECEIVED TO ASSIST WITH THE INVESTIGATION. THERE WAS NO LOT NUMBER PROVIDED AND THEREFORE A BATCH RECORD REVIEW COULD NOT BE PERFORMED. HOWEVER, FUNCTIONAL TESTING WAS PERFORMED AND THE SAMPLE WAS FOUND TO HAVE MET THE MINIMUM AEROSOL OUTPUT REQUIREMENT AT THE LOWER AIR FLOW RATE OF 2.24M. THERE IS NOT ENOUGH INFORMATION TO CONCLUDE THE PRODUCT DID NOT MEET SPECIFICATION AND PERFORM AS INTENDED. NO FURTHER ACTIONS ARE REQUIRED, AND THE COMPLAINT WILL BE CLOSED. NO ADDITIONAL PATIENT/EVENT DETAILS HAVE BEEN PROVIDED TO DATE BUT HAS BEEN REQUESTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. NOTE: THERE ARE FOUR (4) CASES ASSOCIATED WITH THIS PRODUCT COMPLAINT ISSUE; THEREFORE A SEPARATE FDA FORM 3500A HAS BEEN GENERATED TO ADDRESS THE OTHER THREE (3) CASES.

Description of Event or Problem · 1

IT WAS REPORTED BY THE NURSE THAT DURING USE WITH A PATIENT, THE MEDICATION IN THE OPTI-MIST "BUBBLED" IN THE NEBULIZER; HOWEVER, THE TREATMENT MEDICATION DID NOT "ATOMIZE". NO ADDITIONAL PATIENT OR PRODUCT INFORMATION COULD BE OBTAINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
492382 UNO OPTI-MIST+ NEB 210 CM (10/15) INT NEBULIZER (DIRECT PATIETN INTERFACE) CAF UNOMEDICAL S.A. DE C.V. 3772MM

Patients

Seq Age Sex Outcome Treatment
1