FDA Adverse Event Malfunction Summary report: N

AESPIRE 710

MDR report key: 4961365 · Received July 30, 2015

Report

Report Number
9710602-2015-00009
Event Type
Malfunction
Date Received
July 30, 2015
Date of Event
May 21, 2015
Report Date
May 21, 2015
Manufacturer
GE MEDICAL SYSTEMS CHINA CO., LTD.
Product Code
BSZ
PMA / PMN Number
K000706
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS MALFUNCTION WAS DETERMINED TO BE REPORTABLE AS THIS SAME MALFUNCTION HAS PREVIOUSLY CONTRIBUTED TO A SERIOUS INJURY. REFERENCE MDR 2112667-2013-00005. THE DISTRIBUTOR PERFORMED A CHECKOUT OF THE EQUIPMENT AND NOTED THAT THE FLOW SENSOR WAS CRACKED AND THE FLOW SENSOR DIAPHRAGM WAS STUCK TO THE TOP OF THE FLOW SENSOR HOUSING. FLOW SENSORS OF THIS TYPE ARE CUSTOMER REPLACEABLE, ARE RECOMMEND FOR REPLACEMENT AFTER 3 MONTHS, AND ARE WARRANTED FOR 6 MONTHS. THE MAINTENANCE SCHEDULE IN THE USER REFERENCE MANUAL STATES: "REPLACE THE DISPOSABLE FLOW SENSOR (PLASTIC). UNDER TYPICAL USE, THE SENSOR MEETS SPECIFICATIONS FOR A MINIMUM OF 3 MONTHS." IN ENGINEERING EVALUATION, THE STUCK DIAPHRAGM HAS BEEN ABLE TO BE REPRODUCED BY: A HARD IMPACT, SUCH AS DROPPING THE FLOW SENSOR, OR BY; STICKING AN OBJECT INTO THE FLOW SENSOR, CAUSING THE DIAPHRAGM TO STICK OPEN. IF A SENSOR IS SUBJECTED TO A HARD IMPACT, IT IS STILL UNLIKELY THAT THE DIAPHRAGM WILL GET STUCK IN THE OPEN POSITION. THIS FAILURE MODE REQUIRES AN IMPACT IN A VERY LIMITED ORIENTATION TO RESULT IN THE INERTIA NEEDED TO FORCE THE DIAPHRAGM INTO THE STUCK OPEN POSITON. A LEAK CONDITION WILL BE NOTED DURING THE PREOPERATIVE CHECKOUT, AS CONTAINED IN THE AESTIVA USER REFERENCE MANUAL.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE DISTRIBUTOR REPORTED A LEAK WAS NOTED DURING THE PREOPERATIVE CHECKOUT. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
498163 AESPIRE 710 ANESTHESIA GAS MACHINE BSZ GE MEDICAL SYSTEMS CHINA CO., LTD.

Patients

Seq Age Sex Outcome Treatment
1