INTELEPACS
Report
- Report Number
- 9615916-2015-00002
- Event Type
- Malfunction
- Date Received
- July 30, 2015
- Report Date
- July 28, 2015
- Manufacturer
- INTELERAD MEDICAL SYSTEMS INCORPORATED
- Product Code
- LLZ
- PMA / PMN Number
- K083520
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
IN ORDER TO CORRECT THIS ISSUE, INTELERAD WILL: PROVIDE THE AFFECTED CLIENTS WITH A LIST OF STUDIES THAT WERE AFFECTED AND POSSIBLY RECOVER THE MISSING IMAGES. DEPLOY A SOFTWARE UPDATE THAT INCLUDES A FIX AND MITIGATION TO ALL CLIENTS WHERE THE ISSUE COULD POTENTIALLY OCCUR. THIS WILL BE DONE TENTATIVELY. PERIODICALLY AUDIT ALL AFFECTED CLIENT, ANY NEW OCCURRENCES OF THIS ISSUE UNTIL THE MITIGATION (SOFTWARE UPDATE) IS IN PLACE. THE SOFTWARE UPDATE WILL INCLUDE ONLY BUG FIXES - YOUR WORKFLOW WILL NOT BE AFFECTED. THE UPDATE WILL RESULT IN ABOUT 5 MINUTES OF DOWNTIME PER SERVER AS WE PERFORM A FULL SERVICE RESTART. IMPACT ON USERS WILL BE MINIMAL, IF ANY. CORRECTIVE ACTION CALENDAR: FOR ALL CLIENTS WITH PACS VERSION 4-5-1 OR HIGHER, THE SOFTWARE UPDATE WILL BE DEPLOYED WITHIN A TIMEFRAME TO BE AGREED WITH CLIENTS (COMPLETION ESTIMATED BY END OF MARCH 2016).
NO PATIENT INCIDENT REPORTED AT THIS MOMENT - THIS ISSUE IS DOCUMENTED AS A PRECAUTION. FIVE RELATED CUSTOMER COMPLAINTS ((B)(4)) WERE REGISTERED AND INVESTIGATED FROM OCT 2014 TO JUNE 2015 FOR RELATED ISSUES. THE INVESTIGATION OF THESE COMPLAINTS LED TO THE DISCOVERY OF ISSUES WITH IMAGE INGESTION. BUGS IDENTIFIED: (B)(4) LDS: LDS: FAILED TO INGEST IMAGE WITH OVERLAYS BELONGING TO 60XX GROUP HAVING ALPHANUMERIC ID (IE, 600A) (B)(4) LDS: ROWS STUCK IN INCOMING_FILES DUE TO FILES MISSING IN THE INCOMING FILES FOLDER. INVESTIGATION: INVESTIGATION OF THE LOGS IN (B)(4) REVEALED THAT SOME OF THE EXPECTED OVERLAY FILES WERE NOT FOUND ON THE FILE SYSTEM; THIS LED TO DISCOVERY OF THE MISSING FILES LEFT BEHIND AT AN EARLIER STAGE OF IMAGE INGESTION. ROOT CAUSE OF THE PROBLEM: DUE TO THE ASYNCHRONOUS NATURE OF IMAGE INGESTION PROCESSING, A FAILURE AT THE EARLY STAGES OF THE INGESTION PROCESS CAN LEAD TO IMAGES NOT BEING VISIBLE TO THE END USER. THERE IS NO INDICATION TO THE USER THAT THIS HAS OCCURRED. REASON FOR POSSIBLE PATIENT SAFETY: THE RADIOLOGIST MAY PROVIDE A DIAGNOSIS WITHOUT BEING AWARE THAT IMAGES ARE MISSING. DESIGN MITIGATION TO REDUCE THE RISK: PROPOSED NEW MITIGATION: THE PACS WILL REPORT PROBLEMS INGESTING IMAGES AND INDICATE SUCH ISSUES TO THE USER ((B)(4)). THIS MITIGATION WILL BE INCLUDED IN A COMING SOFTWARE RELEASE. (B)(4): LDS: EXPOSE BROKEN SERIES TO USER (WARN ABOUT ADDITIONAL IMAGES PENDING INGESTION WHEN VIEWING IMAGES). DETECTABILITY OF PROBLEM: UNTIL THE SOFTWARE UPDATE THAT WILL INCLUDE THE PROPOSED MITIGATION IS INSTALLED ON THE CLIENT'S SYSTEM, THE USER MAY NOT BE AWARE THAT SOME IMAGES MAY BE MISSING. HOWEVER, IF THE CLIENT'S SYSTEM HAS OTHER MECHANISMS TO VERIFY IMAGE COUNT BETWEEN SOURCE AND DESTINATION, THEN THE USER MIGHT BE ABLE TO KNOW ABOUT THE DISCREPANCY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 495273 | INTELEPACS | PICTURE ARCHIVING AND COMMUNICATION | LLZ | INTELERAD MEDICAL SYSTEMS INCORPORATED | VERSIONS 4.5.1 AND UP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |