BD PARADIGM LINK BLOOD GLUCOSE MONITOR
MDR report key: 495952
·
Received November 6, 2003
Report
- Report Number
- 2243072-2003-00092
- Event Type
- Other
- Date Received
- November 6, 2003
- Date of Event
- October 6, 2003
- Report Date
- November 5, 2003
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER REPORTS GETTING INACCURATE READINGS ON FAMILY MEMBER'S PARADIGM LINK METER. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVERAGE (50) VS. BD READINGS (22, 77, 74) YIELDED DATA POINTS IN THE "A/D" RANGE OF THE GRID - OUTSIDE ACCEPTABLE RANGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD PARADIGM LINK BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | 3064062 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |