RAPIDSTRAND RX
Report
- Report Number
- 2915056-2015-00002
- Event Type
- Other
- Date Received
- July 29, 2015
- Report Date
- July 8, 2015
- Manufacturer
- GE HEALTHCARE
- Product Code
- KXK
- PMA / PMN Number
- K063177
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE STRANDING MATERIAL FOR THE IMPLANTED STRANDS WAS (B)(4) THERASLEEVE. RISK ASSESSMENT: STRANDING OF SEEDS WITH BIO ABSORBABLE SUTURES WAS DEVELOPED 20 YEARS AGO TO REDUCE SEED MIGRATION AND IMPROVE IMPLANT DOSIMETRY. RSRX CAN BE STRANDED WITH TWO TYPES OF STRANDING MATERIAL AT THE CUSTOMER'S REQUEST. A BIO ABSORBABLE BRAIDED SUTURE COMPOSED OF 90/10 (GLYCOLIDE L- LACTIDE) FOR WHICH ABSORPTION OCCURS IN 56-70 DAYS. A BIO ABSORBABLE MONOFILAMENT SLEEVE COMPOSED OF 20/80 (GLYCOLIDE L- LACTIDE) FOR WHICH ABSORPTION OCCURS IN 140 -180 DAYS. THE FIRST TYPE OF STRANDING MATERIAL IS THE SAME AS THAT USED FOR GE/ONCURA'S PRODUCT RAPID STRAND (IMC 7000). IT HAS BEEN IN USE FOR MANY YEARS AND NO CONFIRMED REPORTS OF WHOLE STAND SLIPPAGE HAVE BEEN DOCUMENTED. CONCLUSION: WHOLE STRAND MIGRATION IS AN UNDESIRABLE AND UNUSUAL OCCURRENCE AS A RESULT OF PROSTATE LOW DOSE RATE SEED IMPLANTATION FOR LOCALISED PROSTATE CANCER. IF NOT IDENTIFIED THE ABSENCE OR CHANGE OF LOCATION OF SEEDS IN THE PROSTATE TISSUE DUE TO MIGRATION MAY CAUSE A COLD SPOT WHICH COULD COMPROMISE THE EFFICACY OF THE TREATMENT. ADDITIONALLY THE PRESENCE OF SEEDS IN UNPLANNED AREAS OF THE PROSTATE OR NEIGHBORING STRUCTURES MAY GIVE RISE TO UNEXPECTED MORBIDITY.
EVALUATION SUMMARY: CONCLUSION: WHOLE STRAND MIGRATION IS AN UNDESIRABLE AND UNUSUAL OCCURRENCE AS A RESULT OF PROSTATE LOW DOSE RATE SEED IMPLANTATION FOR LOCALISED PROSTATE CANCER. IF NOT IDENTIFIED THE ABSENCE OR CHANGE OF LOCATION OF SEEDS IN THE PROSTATE TISSUE DUE TO MIGRATION MAY CAUSE A COLD SPOT WHICH COULD COMPROMISE THE EFFICACY OF THE TREATMENT. ADDITIONALLY, THE PRESENCE OF SEEDS IN UNPLANNED AREAS OF THE PROSTATE OR NEIGHBOURING STRUCTURES MAY GIVE RISE TO UNEXPECTED MORBIDITY.
REPORT # 2915056-2015-00002 IS A PHYSICIAN REPORT FROM (B)(6) CONCERNING A DEVICE MALFUNCTION WITH NO ASSOCIATED ADVERSE EVENT. THE REPORT INVOLVES A MALE PATIENT (AGE NOT REPORTED) WHO EXPERIENCED SIGNIFICANT DISLOCATION FROM THE ORIGINAL IMPLANTED POSITION OF RAPID STRAND RX (RSRX) FOR THE LOW DOSE RATE BRACHYTHERAPY TREATMENT OF LOCALISED PROSTATE CANCER. CONCURRENT MEDICAL CONDITION INCLUDED PROSTATE CANCER. PAST MEDICAL HISTORY WAS NOT PROVIDED. CONCOMITANT MEDICATION WAS NOT PROVIDED. ON AN UNKNOWN DATE, THE PATIENT UNDERWENT AN IMPLANTATION OF RAPIDSTRAND RX FOR THE LOW DOSE RATE BRACHYTHERAPY TREATMENT OF LOCALISED PROSTATE CANCER (DOSE NOT REPORTED). THE LOT NUMBER WAS PROVIDED AS 0300134679. THE INTRAOPERATIVE FLUOROSCOPIC IMAGES SHOWED IMPLANTS WITH A SATISFACTORY DISTRIBUTION OF SEEDS. HOWEVER THE POST IMPLANT DOSIMETRY IMAGES SHOWED SIGNIFICANT DISLOCATION OF SEEDS FROM EXPECTED POSITIONS WHICH LED THE ATTENDING PHYSICIAN TO CONCLUDE THAT WHOLE STRANDS HAD MIGRATED AT AN UNDEFINED TIME AFTER THE IMPLANTATION. MEDICAL ASSESSMENT CONCLUDED: WHOLE STRAND MIGRATION IS AN UNDESIRABLE AND UNUSUAL OCCURRENCE AS A RESULT OF PROSTATE LOW DOSE RATE SEED IMPLANTATION FOR LOCALISED PROSTATE CANCER. IF NOT IDENTIFIED THE ABSENCE OR CHANGE OF LOCATION OF SEEDS IN THE PROSTATE TISSUE DUE TO MIGRATION MAY CAUSE A COLD SPOT WHICH COULD COMPROMISE THE EFFICACY OF THE TREATMENT. ADDITIONALLY THE PRESENCE OF SEEDS IN UNPLANNED AREAS OF THE PROSTATE OR NEIGHBORING STRUCTURES MAY GIVE RISE TO UNEXPECTED MORBIDITY. OUTCOME OF THE EVENT WAS NOT PROVIDED AT THE TIME OF THIS REPORT. QUALITY ASSURANCE INVESTIGATION HAS BEEN INITIATED.
FOLLOW-UP INFORMATION RECEIVED ON 09/14/2015: QUALITY ASSURANCE INVESTIGATION SUMMARY: ALL DOCUMENTS AND ALL IN PROCESS PAPERWORK WERE REVIEWED FOR TO DETERMINE IF ANY ISSUES OCCURRED DURING THE PRODUCTION AND LOADING PROCESS. THE DIGITAL IMAGES OF THE NEEDLE TRAYS SHOW THAT THE SEEDS AND SPACERS WERE BUILT PER ORDER SPECIFICATIONS IN ACCORDANCE TO THE CUSTOMER NEEDLE CONFIGURATION PLANS. INSPECTION RECORDS FOR THE STRAND MATERIAL WERE REVIEWED AND APPROVED PRIOR TO THE LOADING PROCESS. THE ORDER MET ALL RELEASE SPECIFICATIONS AND REQUIREMENTS AND THERE WERE NO MANUFACTURING ANOMALIES OR UNUSUAL EVENTS DURING MANUFACTURING. QUALITY ASSURANCE INVESTIGATION FINALIZED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 495268 | RAPIDSTRAND RX | RADIONUCLIDE BRACHYTHERAPY SOURCE | KXK | GE HEALTHCARE | 6711 | 0300134679 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |