G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2015-65153
- Event Type
- Malfunction
- Date Received
- July 30, 2015
- Date of Event
- July 6, 2015
- Report Date
- July 6, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4).
THE COMPLAINT SENSOR WAS NOT RETURNED TO DEXCOM. THE RECEIVER (PART NUMBER STK-GL-113/SERIAL NUMBER (B)(4)/LOT NUMBER 5200547), BEING USED WITH THE COMPLAINT SENSOR, WAS RETURNED ON (B)(6)2015. THE RETURNED RECEIVER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION; HOWEVER, A REVIEW OF THE DIAGNOSTIC CHART CONFIRMED THE REPORTED EVENT OF INACCURACIES. A ROOT CAUSE COULD NOT BE DETERMINED.
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT CONTINUOUS GLUCOSE MONITOR (CGM) INACCURACY COMPARED TO BLOOD GLUCOSE (BG) METER ON (B)(6) 2015. SENSOR WAS INSERTED ON (B)(6) 2015. PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 496319 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9500-25 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |