ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT
Report
- Report Number
- 1419322-2015-00068
- Event Type
- Malfunction
- Date Received
- July 24, 2015
- Report Date
- June 24, 2015
- Manufacturer
- DENTSPLY PROFESSIONAL
- Product Code
- EKX
- PMA / PMN Number
- K031145
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- DENTIST
Narratives
THOUGH NO MEDICAL/SURGICAL INTERVENTION WAS REQUIRED TO PRECLUDE A SERIOUS INJURY IN THIS EVENT, THERE HAVE BEEN PREVIOUSLY REPORTED EVENTS INVOLVING THIS DEVICE THAT RESULTED IN THE NEED FOR MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION. THEREFORE, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE COMPLAINT WAS VERIFIED THROUGH PRODUCTION AND QUALITY TESTING. MAXIMUM TEMPERATURE FOR THE ATTACHMENT HEAD EXCEEDED ISO 13732-1 AND ES-1104 FOR MAXIMUM ALLOWABLE TEMPERATURE DURING TESTING. CAP END BEARING RETAINER FAILURE, FREE ROAMING BALL BEARINGS AND EXCESSIVE DEBRIS MOST LIKELY CREATED FRICTION WITHIN THE HEAD WHICH RESULTED IN TEMPERATURE INCREASE. GEAR FUNCTION WAS ALSO COMPROMISED INSIDE THE HEAD AND NECK WHICH IS EVIDENT FROM THE WEAR ON THE TEETH OF THE GEARS.
IN THIS EVENT A DOCTOR REPORTED THAT AN ESTYLUS 1:5 HANDPIECE OVERHEATED WHILE IN USE. THERE WAS NO INJURY OR INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 482203 | ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT | EKX | DENTSPLY PROFESSIONAL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |