ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2015-26233
- Event Type
- Malfunction
- Date Received
- July 30, 2015
- Report Date
- July 15, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/15/2015 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PCB BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED. EVALUATION ALSO REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED BELOW BUMPER PAD. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).
THE PUMP WAS RETURNED FOR INVESTIGATION. EVALUATION REVEALED A BATTERY COMPARTMENT CRACK. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2015.