ELECTRODE, CUTTING LOOP, 26FR
Report
- Report Number
- 1418479-2015-00020
- Event Type
- Malfunction
- Date Received
- July 27, 2015
- Report Date
- June 29, 2015
- Manufacturer
- RICHARD WOLF MEDICAL INSTRUMENTS CORP.
- Product Code
- FAS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- NURSE
Narratives
AN INVESTIGATION COULD NOT BE COMPLETED AS THE ACTUAL DEVICE WAS NOT RETURNED TO THE RWMIC FACILITY AS OF 07/24/2015. A REQUEST FOR ADD'L INFO AS WELL AS PHOTOS OF DEVICE SUBMITTED. NO RESPONSE AS OF 07/24/2015. DEVICE MFR'D FEB 2015. THE TWO MOST COMMON SCENARIOS ON HOW TIP COULD HAVE BROKEN OFF IS AS FOLLOWS: TIP BENT; EXCESSIVE FORCE APPLIED OR TIP CAME INTO CONTACT W/HARD OBJECT., TIP OVER HEATED; EXTENDED USE OF DEVICE OR ACTIVATED WHILE OUTSIDE OF LIQUID. NOT UNCOMMON FOR TIP TO DISAPPEAR OR VAPORIZE WHEN THIS SCENARIO OCCURS. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. I E INTENDED USE, INDICATIONS AND FIELD OF USE, PREPARATION AND CAUTIONS REGARDING HIGH TEMPERATURES AND USE OF IRRIGATION FLUID. NO SIMILAR ISSUES HAS OCCURRED ON THIS DEVICE IN THE LAST THREE YEARS. RWMIC CONSIDERS THIS MATTER CLOSED. HOWEVER, IN THE EVENT RWMIC RECEIVE'S ADD'L INFO, A FOLLOW-UP REPORT W/BE PROVIDED TO FDA.
RICHARD WOLF MED. INSTRUMENTS CORP. (RWMIC) WAS RECENTLY NOTIFIED BY FACILITY WHICH REPORTED A PORTION OF DEVICE BROKE OFF AND FELL INTO PATIENT. THE FOREIGN OBJECT WAS EASILY RETRIEVED AND PROCEDURE COMPLETED AS SCHEDULED. NO INJURY TO PATIENT OR STAFF REPORTED. TIP OF DEVICE BROKE OFF AND FACILITY SWAPPED OUT TO BACK UP DEVICE, SAME DEVICE ID AND LOT NUMBER. TIP OF BACK UP DEVICE ALSO BROKE OFF. PROCEDURE COMPLETED W/THIRD DEVICE. TWO MDRS WILL BE SUBMITTED FOR THIS EVENT. REF MDR #1418479-2015-00021.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 485531 | ELECTRODE, CUTTING LOOP, 26FR | FAS | RICHARD WOLF MEDICAL INSTRUMENTS CORP. | 8426.1420 | 51002653 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |