TRUFREEZE CRYOSPRAY SYSTEM
Report
- Report Number
- 3010140265-2015-00007
- Event Type
- Malfunction
- Date Received
- July 30, 2015
- Date of Event
- July 15, 2015
- Report Date
- July 30, 2015
- Manufacturer
- CSA MEDICAL INC.
- Product Code
- GEH
- PMA / PMN Number
- K152668
- Removal / Correction Number
- 9062377-11/10/14-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION CONCLUSION - COMPONENT FAILURE (SOUND DRIVER).
THE TRUFREEZE CRYOTHERAPY SYSTEM PROVIDES A TIMER FOR THE USER TO TIME CRYOSPRAY TREATMENTS. THE TIMER DOES NOT START OR STOP CRYOSPRAYS. THE TIMER IS DISPLAYED ON THE PANEL PC'S TIMER PANEL IN MINUTES:SECONDS (0:00) FORMAT ALONG WITH A TIMER PROGRESS BAR. AUDIBLE TONES ARE GENERATED IN CONJUNCTION WITH THE PROGRESS BAR. THE TONES ARE GENERATED AT FIVE-SECOND INTERVALS AND AT EACH SECOND OF THE FINAL FIVE SECONDS APPROACHING THE TARGET SET TIME WHEN THE CRYOGEN PEDAL IS ENGAGED. IN ADDITION, USERS ARE TRAINED TO HAVE A NURSE OR TECHNICIAN NOTIFY THE TREATING PHYSICIAN WHEN THE SPRAY TIME HAS BEEN COMPLETED. A LOSS OF THE AUDIBLE TONES HAS BEEN REPORTED ON SYSTEMS CONFIGURED WITH THE ADVANTECH POC-127 PANEL PC ON LIMITED OCCASIONS WHEN THERE HAS BEEN NO MANUAL INPUT INTO THE PANEL PC FOR EXTENDED PERIODS OF TIME ( >10 MINUTES). THE TIMER DISPLAY FUNCTIONS CONTINUE TO OPERATE AS INTENDED. THERE HAVE BEEN NO ASSOCIATED ADVERSE EVENTS RELATED TO THIS ISSUE. HOWEVER, THERE IS THE POSSIBILITY THAT MORE CRYOGEN COULD BE DELIVERED THAN EXPECTED BY THE USER IF THEY RELY SOLELY ON THE AUDIBLE TONE. IN A PASSIVE VENTING PROCEDURE, THIS POTENTIAL EXCESS CRYOGEN DELIVERY COULD INDUCE BRADYCARDIA. THE INSTRUCTIONS FOR USE HAS BEEN UPDATED TO INCLUDE "THE PHYSICIAN SHOULD KEEP TRACK OF THE DURATION OF DELIVERY BY UTILIZING THE TIMER'S AUDIBLE TONE AND BY ASSIGNING AN APPROPRIATE HEALTH CARE PROVIDER TO AUDIBLY CALL OUT THE TIMER'S SECOND BY SECOND DISPLAY UNTIL THE SELECTED LENGTH OF CRYOGEN SPRAY HAS COMPLETED." THIS PROVIDES A MITIGATION TO ADDRESS THE RARE INSTANCE OF LOSS OF AUDIBLE TONE.
THE AUDIBLE SOUND ASSOCIATED WITH THE TIMER WAS LOST DURING A PASSIVE VENTING (AIRWAY) CASE. THERE WAS NO REPORTED ADVERSE EVENT.
THE AUDIBLE SOUND ASSOCIATED WITH THE TIMER WAS LOST DURING A PASSIVE VENTING (AIRWAY) CASE. THERE WAS NO REPORTED ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 498817 | TRUFREEZE CRYOSPRAY SYSTEM | UNIT, CRYOSURGICAL, ACCESSORIES | GEH | CSA MEDICAL INC. | CC3-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |