FDA Adverse Event Malfunction Summary report: N

TRUFREEZE CRYOSPRAY SYSTEM

MDR report key: 4951967 · Received July 30, 2015

Report

Report Number
3010140265-2015-00007
Event Type
Malfunction
Date Received
July 30, 2015
Date of Event
July 15, 2015
Report Date
July 30, 2015
Manufacturer
CSA MEDICAL INC.
Product Code
GEH
PMA / PMN Number
K152668
Removal / Correction Number
9062377-11/10/14-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION CONCLUSION - COMPONENT FAILURE (SOUND DRIVER).

Additional Manufacturer Narrative · 1

THE TRUFREEZE CRYOTHERAPY SYSTEM PROVIDES A TIMER FOR THE USER TO TIME CRYOSPRAY TREATMENTS. THE TIMER DOES NOT START OR STOP CRYOSPRAYS. THE TIMER IS DISPLAYED ON THE PANEL PC'S TIMER PANEL IN MINUTES:SECONDS (0:00) FORMAT ALONG WITH A TIMER PROGRESS BAR. AUDIBLE TONES ARE GENERATED IN CONJUNCTION WITH THE PROGRESS BAR. THE TONES ARE GENERATED AT FIVE-SECOND INTERVALS AND AT EACH SECOND OF THE FINAL FIVE SECONDS APPROACHING THE TARGET SET TIME WHEN THE CRYOGEN PEDAL IS ENGAGED. IN ADDITION, USERS ARE TRAINED TO HAVE A NURSE OR TECHNICIAN NOTIFY THE TREATING PHYSICIAN WHEN THE SPRAY TIME HAS BEEN COMPLETED. A LOSS OF THE AUDIBLE TONES HAS BEEN REPORTED ON SYSTEMS CONFIGURED WITH THE ADVANTECH POC-127 PANEL PC ON LIMITED OCCASIONS WHEN THERE HAS BEEN NO MANUAL INPUT INTO THE PANEL PC FOR EXTENDED PERIODS OF TIME ( >10 MINUTES). THE TIMER DISPLAY FUNCTIONS CONTINUE TO OPERATE AS INTENDED. THERE HAVE BEEN NO ASSOCIATED ADVERSE EVENTS RELATED TO THIS ISSUE. HOWEVER, THERE IS THE POSSIBILITY THAT MORE CRYOGEN COULD BE DELIVERED THAN EXPECTED BY THE USER IF THEY RELY SOLELY ON THE AUDIBLE TONE. IN A PASSIVE VENTING PROCEDURE, THIS POTENTIAL EXCESS CRYOGEN DELIVERY COULD INDUCE BRADYCARDIA. THE INSTRUCTIONS FOR USE HAS BEEN UPDATED TO INCLUDE "THE PHYSICIAN SHOULD KEEP TRACK OF THE DURATION OF DELIVERY BY UTILIZING THE TIMER'S AUDIBLE TONE AND BY ASSIGNING AN APPROPRIATE HEALTH CARE PROVIDER TO AUDIBLY CALL OUT THE TIMER'S SECOND BY SECOND DISPLAY UNTIL THE SELECTED LENGTH OF CRYOGEN SPRAY HAS COMPLETED." THIS PROVIDES A MITIGATION TO ADDRESS THE RARE INSTANCE OF LOSS OF AUDIBLE TONE.

Description of Event or Problem · 1

THE AUDIBLE SOUND ASSOCIATED WITH THE TIMER WAS LOST DURING A PASSIVE VENTING (AIRWAY) CASE. THERE WAS NO REPORTED ADVERSE EVENT.

Description of Event or Problem · 1

THE AUDIBLE SOUND ASSOCIATED WITH THE TIMER WAS LOST DURING A PASSIVE VENTING (AIRWAY) CASE. THERE WAS NO REPORTED ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
498817 TRUFREEZE CRYOSPRAY SYSTEM UNIT, CRYOSURGICAL, ACCESSORIES GEH CSA MEDICAL INC. CC3-01

Patients

Seq Age Sex Outcome Treatment
1