FDA Adverse Event Injury Summary report: N

INTERA 1.5T ACHIEVA PULSAR

MDR report key: 4949115 · Received July 29, 2015

Report

Report Number
3003768277-2015-00061
Event Type
Injury
Date Received
July 29, 2015
Date of Event
July 23, 2015
Report Date
July 24, 2015
Manufacturer
PHILIPS HEALTHCARE
Product Code
LNH
PMA / PMN Number
K030520
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
RADIOLOGIC TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

BASED ON THE PROVIDED INFORMATION IT IS CONCLUDED THAT THE INJURY AS SUSTAINED BY THE PATIENT IS CAUSED BY A USER ERROR. DURING MOVEMENT OF THE TABLE INTO THE BORE THE PATIENT SUDDENLY GRABBED HOLD OF THE EDGE OF THE TABLE TOP. THIS LED TO THE RIGHT RING FINGER OF THE PATIENT BEING PINCHED BETWEEN THE TABLETOP AND THE BORE COVERS. WITHIN THE INSTRUCTIONS FOR USE THERE ARE WARNINGS RELATED TO MOVING THE PATIENT INTO THE BORE. DURING MOVEMENT THE OPERATOR SHOULD ALWAYS CHECK THE PATIENT'S HANDS AND MAKE SURE THERE ARE NO CABLES OR EXTREMITIES POSITIONED SUCH THAT THEY CAN BE CAUGHT. (B)(4).

Description of Event or Problem · 1

PHILIPS RECEIVED A REPORT FROM A CUSTOMER THAT A PATIENT INJURED THE RING FINGER OF THE RIGHT HAND AFTER BEING MOVED INTO THE BORE OF THE MR SYSTEM. THE PATIENT'S RING FINGER GOT PINCHED BETWEEN THE MAGNET COVER AND TABLE TOP. THE PATIENT'S RIGHT RING FINGER WAS CUT AND SIX STITCHES WERE NEEDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
494435 INTERA 1.5T ACHIEVA PULSAR SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING LNH PHILIPS HEALTHCARE 781171

Patients

Seq Age Sex Outcome Treatment
1 83 YR Other