INTERA 1.5T ACHIEVA PULSAR
Report
- Report Number
- 3003768277-2015-00061
- Event Type
- Injury
- Date Received
- July 29, 2015
- Date of Event
- July 23, 2015
- Report Date
- July 24, 2015
- Manufacturer
- PHILIPS HEALTHCARE
- Product Code
- LNH
- PMA / PMN Number
- K030520
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- RADIOLOGIC TECHNOLOGIST
Narratives
BASED ON THE PROVIDED INFORMATION IT IS CONCLUDED THAT THE INJURY AS SUSTAINED BY THE PATIENT IS CAUSED BY A USER ERROR. DURING MOVEMENT OF THE TABLE INTO THE BORE THE PATIENT SUDDENLY GRABBED HOLD OF THE EDGE OF THE TABLE TOP. THIS LED TO THE RIGHT RING FINGER OF THE PATIENT BEING PINCHED BETWEEN THE TABLETOP AND THE BORE COVERS. WITHIN THE INSTRUCTIONS FOR USE THERE ARE WARNINGS RELATED TO MOVING THE PATIENT INTO THE BORE. DURING MOVEMENT THE OPERATOR SHOULD ALWAYS CHECK THE PATIENT'S HANDS AND MAKE SURE THERE ARE NO CABLES OR EXTREMITIES POSITIONED SUCH THAT THEY CAN BE CAUGHT. (B)(4).
PHILIPS RECEIVED A REPORT FROM A CUSTOMER THAT A PATIENT INJURED THE RING FINGER OF THE RIGHT HAND AFTER BEING MOVED INTO THE BORE OF THE MR SYSTEM. THE PATIENT'S RING FINGER GOT PINCHED BETWEEN THE MAGNET COVER AND TABLE TOP. THE PATIENT'S RIGHT RING FINGER WAS CUT AND SIX STITCHES WERE NEEDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 494435 | INTERA 1.5T ACHIEVA PULSAR | SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING | LNH | PHILIPS HEALTHCARE | 781171 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Other |