FDA Adverse Event Injury Summary report: N

SOL SYS CEM PLUG DRILLS 4.5MM

MDR report key: 4946966 · Received July 28, 2015

Report

Report Number
1818910-2015-26753
Event Type
Injury
Date Received
July 28, 2015
Date of Event
July 23, 2015
Report Date
July 23, 2015
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
LXH
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. (B)(4). THIS COMPLAINT IS THE SUBJECT OF LITIGATION OR A LEGAL CLAIM AND CURRENTLY COMPLETE PRODUCT DETAIL IS NOT AVAILABLE AT THIS TIME. A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE

Additional Manufacturer Narrative · 1

CONCLUSION AND JUSTIFICATION STATUS FOR MDR: EXAMINATION OF THE RETURNED DRILL (PRODUCT CODE 2108-08-000) CONFIRMED THE TIP BROKE OFF. IT WAS DETERMINED THAT THE CEMENT PLUG DRILL WAS CYCLICALLY LOADED BEYOND THE MATERIAL LIMIT RESULTING IN A HIGH STRESS LOW CYCLE FATIGUE CRACK INITIATION, THEN QUICKLY OVER LOAD FRACTURED DUE TO A MIXED MODE OF DUCTILE AND BRITTLE OVERLOAD OF THE MATERIAL. NO EVIDENCE OF MATERIAL OR MANUFACTURING DEFECTS WAS OBSERVED. THE DRILL, PRODUCT CODE 2108-04-000, WAS NOT RETURNED THEREFORE EXAMINATION WAS NOT POSSIBLE. INITIAL REPORTING STATED THE SOLUTION CEMENT PLUG DRILL 4.5MM WOULD NOT BE RETURNED. A SEARCH OF THE COMPLAINT DATABASE WAS NOT POSSIBLE SINCE THE LOT NUMBER WAS NOT PROVIDED. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. COMPLAINTS WILL BE MONITORED THROUGH POST MARKET SURVEILLANCE (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

WHILE REMOVING A STRYKER STEM, THE CEMENT DRILL BROKE IN THE FEMUR. LATER IN THE SURGERY, THE SECOND DRILL BROKE. THE SURGEON WAS ABLE TO REMOVE THE BROKEN PIECES. THE NEW STEM WAS THEN INSERTED WITHOUT ANY FURTHER ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
489175 SOL SYS CEM PLUG DRILLS 4.5MM HIPS INSTRUMENTS LXH DEPUY ORTHOPAEDICS, INC. 1818910 UNK

Patients

Seq Age Sex Outcome Treatment
1 73 YR Other