30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT
Report
- Report Number
- 2424472-2015-00042
- Event Type
- Malfunction
- Date Received
- July 23, 2015
- Report Date
- June 23, 2015
- Manufacturer
- DENTSPLY PROFESSIONAL
- Product Code
- ELC
- PMA / PMN Number
- K052334
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- DENTIST
Narratives
WHILE NO SERIOUS INJURY RESULTED IN THIS EVENT, THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH RESULTS ARE NOT AVAILABLE AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
INSERT WAS EVALUATED AND THE HOTTEST OBSERVED TEMPERATURE WAS 92.4F. THIS WAS WELL WITHIN THE LIMITS FOR SAFE USE. A DHR REVIEW WAS CONDUCTED WITH NO DISCREPANCIES NOTED.
IN THIS EVENT, IT WAS REPORTED THAT A CAVITRON 30K FSI-SLI-10S INSERT TIP OVERHEATED; NO INJURY RESULTED.