FDA Adverse Event Malfunction Summary report: N

30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT

MDR report key: 4943580 · Received July 23, 2015

Report

Report Number
2424472-2015-00042
Event Type
Malfunction
Date Received
July 23, 2015
Report Date
June 23, 2015
Manufacturer
DENTSPLY PROFESSIONAL
Product Code
ELC
PMA / PMN Number
K052334
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

WHILE NO SERIOUS INJURY RESULTED IN THIS EVENT, THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH RESULTS ARE NOT AVAILABLE AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

INSERT WAS EVALUATED AND THE HOTTEST OBSERVED TEMPERATURE WAS 92.4F. THIS WAS WELL WITHIN THE LIMITS FOR SAFE USE. A DHR REVIEW WAS CONDUCTED WITH NO DISCREPANCIES NOTED.

Description of Event or Problem · 1

IN THIS EVENT, IT WAS REPORTED THAT A CAVITRON 30K FSI-SLI-10S INSERT TIP OVERHEATED; NO INJURY RESULTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
481271 30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT ULTRASONIC SCALER ELC DENTSPLY PROFESSIONAL 5119

Patients

Seq Age Sex Outcome Treatment
1