CENTURION EXCEL
Report
- Report Number
- 2028295-2015-00001
- Event Type
- Malfunction
- Date Received
- July 22, 2015
- Date of Event
- June 12, 2012
- Report Date
- July 21, 2015
- Manufacturer
- MEDICAL ILLUMINATION INTERNATIONAL, INC.
- Product Code
- LBI
- PMA / PMN Number
- K010495
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE LIGHT AND ARM WERE RETURNED TO THE FACTORY FOR ANALYSIS. UPON ARRIVAL, THE ARM SUPPORT FITTING WAS INSPECTED FOR POROSITY AND/OR VOIDS IN THE MATERIAL. NO ANOMALIES IN THE MATERIAL WERE FOUND. THE ARM SUPPORT MATERIAL WAS ALSO IDENTIFIED THAT IT WAS THE SPECIFIED MATERIAL. ADDITIONAL TESTING OF NEW PRODUCT WAS PERFORMED TO VERIFY PRODUCTION PARTS PERFORMED AS DESIGNED. UNABLE TO DETERMINE EXACT CAUSE OF THIS FAILURE. WILL MONITOR FOR ANY FUTURE OCCURRENCES. THE ARM SUPPORT COMPONENT WAS PUT UNDER TEST WITH A 6X LOAD TO CONFIRM THE INITIAL LOAD HANDLING CAPACITY. THIS TEST WAS CONFIRMED AS POSITIVE AND THE CASE WAS CLOSE WITH THE REPLACEMENT OF THE LIGHT ON (B)(4) 2012.
ON (B)(6) 2012 AN INCIDENT WAS REPORTED TO MEDICAL ILLUMINATION REGARDING AN EXCEL LIGHT ARM SUPPORT FITTING FRACTURED CAUSING THE LIGHT TO FALL HITTING THE DOCTOR ON HIS BACK. THE LIGHT FELL WHEN BEING REMOVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 477103 | CENTURION EXCEL | MINOR SURGICAL LIGHT | LBI | MEDICAL ILLUMINATION INTERNATIONAL, INC. | 061514 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |