FDA Adverse Event Malfunction Summary report: N

CENTURION EXCEL

MDR report key: 4943577 · Received July 22, 2015

Report

Report Number
2028295-2015-00001
Event Type
Malfunction
Date Received
July 22, 2015
Date of Event
June 12, 2012
Report Date
July 21, 2015
Manufacturer
MEDICAL ILLUMINATION INTERNATIONAL, INC.
Product Code
LBI
PMA / PMN Number
K010495
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE LIGHT AND ARM WERE RETURNED TO THE FACTORY FOR ANALYSIS. UPON ARRIVAL, THE ARM SUPPORT FITTING WAS INSPECTED FOR POROSITY AND/OR VOIDS IN THE MATERIAL. NO ANOMALIES IN THE MATERIAL WERE FOUND. THE ARM SUPPORT MATERIAL WAS ALSO IDENTIFIED THAT IT WAS THE SPECIFIED MATERIAL. ADDITIONAL TESTING OF NEW PRODUCT WAS PERFORMED TO VERIFY PRODUCTION PARTS PERFORMED AS DESIGNED. UNABLE TO DETERMINE EXACT CAUSE OF THIS FAILURE. WILL MONITOR FOR ANY FUTURE OCCURRENCES. THE ARM SUPPORT COMPONENT WAS PUT UNDER TEST WITH A 6X LOAD TO CONFIRM THE INITIAL LOAD HANDLING CAPACITY. THIS TEST WAS CONFIRMED AS POSITIVE AND THE CASE WAS CLOSE WITH THE REPLACEMENT OF THE LIGHT ON (B)(4) 2012.

Description of Event or Problem · 1

ON (B)(6) 2012 AN INCIDENT WAS REPORTED TO MEDICAL ILLUMINATION REGARDING AN EXCEL LIGHT ARM SUPPORT FITTING FRACTURED CAUSING THE LIGHT TO FALL HITTING THE DOCTOR ON HIS BACK. THE LIGHT FELL WHEN BEING REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
477103 CENTURION EXCEL MINOR SURGICAL LIGHT LBI MEDICAL ILLUMINATION INTERNATIONAL, INC. 061514

Patients

Seq Age Sex Outcome Treatment
1