FDA Adverse Event Malfunction Summary report: N

PURELY YOURS

MDR report key: 4943568 · Received July 22, 2015

Report

Report Number
3009974348-2015-00155
Event Type
Malfunction
Date Received
July 22, 2015
Date of Event
June 9, 2014
Report Date
June 9, 2014
Manufacturer
AMEDA, INC.
Product Code
HGX
PMA / PMN Number
K973501
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT PURELY YOURS BREAST PUMP WITH AC ADAPTER WAS SENT TO CUSTOMER ON 06/09/2014. CUSTOMER WAS SENT A (B)(6) RETURN LABEL FOR SHIPMENT BACK TO AMEDA, INC. FOR INVESTIGATION OF COMPLAINT. PURELY YOURS MOTOR BASE AND AC ADAPTER WERE NOT RETURNED DESPITE CONTACTING THE CUSTOMER SEVERAL TIMES.

Description of Event or Problem · 1

AS PART OF AMEDA, INC.'S QUALITY MANAGEMENT SYSTEM ACTIVITIES, A REVIEW OF HISTORICAL COMPLAINTS WAS CONDUCTED. IT WAS DETERMINED THAT THIS COMPLAINT SHOULD HAVE BEEN REPORTED TO FDA. CUSTOMER CONTACTED AMEDA, INC. ON 06/09/2014 TO REPORT THE PURELY YOURS BREAST PUMP SHE USES EXCLUSIVELY TO EXPRESS HER BREAST MILK FOR HER INFANT FOR THE PAST 6 MONTHS, ELICITED A BURNING SMELL, MOTOR BASE BECAME HOT AND SHE NOTED SMOKE COMING FROM THE AC ADAPTER NEAR THE PLUG. CUSTOMER REPORTS NO INJURY OR BURN OCCURRED DURING THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
474303 PURELY YOURS ELECTRIC BREAST PUMP, 884.5160 HGX AMEDA, INC. 24502082

Patients

Seq Age Sex Outcome Treatment
1