LMA PROSEAL, REU, SIZE 4
Report
- Report Number
- 9681900-2015-00047
- Event Type
- Malfunction
- Date Received
- July 27, 2015
- Date of Event
- March 19, 2015
- Report Date
- July 3, 2015
- Manufacturer
- THE LARYNGEAL MASK COMPANY
- Product Code
- BTR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED AND THERE WERE NO ISSUES FOUND THAT COULD HAVE CONTRIBUTED TO THE REPORTED FAILURE. ALL PROCESSES WERE EXECUTED ACCORDING TO THE STANDARD OPERATING METHODS. THE DEVICE WAS OBSERVED TO HAVE A BROKEN CONNECTOR (BROKEN INTO 2 PORTIONS WITH 1 PORTION STILL ATTACHED TO AIRWAY TUBE) AND THE AIRWAY TUBE WAS DISCOLORED. THE CONNECTOR WAS OBSERVED TO HAVE CRAZED LINES WHICH WERE SMOOTH TO TOUCH DURING THE INSPECTION. THE REPORTED DEFECT WAS CONFIRMED THROUGH VISUAL INSPECTION. DUE TO THE PROFILE OF THE CRAZED LINES ON THE BROKEN CONNECTOR, THE FAILURE WAS MOST LIKELY DUE TO A CHEMICAL THAT WAS IN CONTACT WITH THE DEVICE. THE STURDINESS OF THE POLYSULPHONE, THE MATERIAL THE CONNECTOR IS MADE OF, WAS IMPACTED. OTHER REMARKS: THE NON-COMPATIBLE DETERGENT, CONCENTRATED DETERGENT OR HIGH AUTOCLAVING TEMPERATURE COULD HAVE ACCELERATED THE DETERIORATION OF THE CONNECTOR. THE CONNECTOR MUST BE INSPECTED FOR CRAZING PRIOR TO USE. THERE IS ALSO A PREVIOUS STUDY CONFIRMING THAT IF THE LMA PROSEAL IS STERILIZED WITH LUBRICATED MEDICAL INSTRUMENTS, THE LUBRICANT WILL ACCELERATE DETERIORATION OF THE CONNECTOR. THE LMA PROSEAL SHOULD BE PROCESSED WITH A MILD DETERGENT SUCH AS ENDOZIME AND AUTOCLAVED (HOLD TIME) 10 MINUTES AT 134 DEGREES CELSIUS.
(B)(4). THE DEVICE SAMPLE WAS NOT RETURNED FOR EVALUATION AT THE TIME OF THIS REPORT.
THE EVENT IS REPORTED AS: THE CUSTOMER ALLEGES THE 15MM CONNECTOR SNAPPED DURING USE. THERE WAS NO REPORTED PATIENT HARM.
THE EVENT IS REPORTED AS:THE CUSTOMER ALLEGES THE 15MM CONNECTOR SNAPPED DURING USE.THERE WAS NO REPORTED PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 486909 | LMA PROSEAL, REU, SIZE 4 | LARYNGEAL MASK AIRWAY | BTR | THE LARYNGEAL MASK COMPANY |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |