FDA Adverse Event Injury Summary report: N

UNKNOWN DEPUY ATTUNE TIBIA COMPONENT

MDR report key: 4942955 · Received July 27, 2015

Report

Report Number
1818910-2015-26539
Event Type
Injury
Date Received
July 27, 2015
Date of Event
January 30, 2015
Report Date
July 9, 2015
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
JWH
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. DEPUY SYNTHES HAS BEEN INFORMED THAT THE CATALOG NUMBER AND LOT NUMBER IS NOT AVAILABLE. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE LOT SPECIFIC COMPLAINT DATABASE SEARCH, OR DEVICE HISTORY RECORD (DHR) REVIEW WAS NOT POSSIBLE BECAUSE THE REQUIRED LOT CODE(S) WAS NOT PROVIDED. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. (B)(4) HAS BEEN UNDERTAKEN TO INVESTIGATE FURTHER. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

PATIENT WAS REVISED DUE TO PAIN AND LOOSENING. DEPUY CEMENT WAS USED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
486411 UNKNOWN DEPUY ATTUNE TIBIA COMPONENT KNEE TIBIAL BEARING/INSERT JWH DEPUY ORTHOPAEDICS, INC. UNK

Patients

Seq Age Sex Outcome Treatment
1 85 YR Required Intervention