RAPIDPOINT 500
Report
- Report Number
- 1217157-2015-00107
- Event Type
- Malfunction
- Date Received
- July 27, 2015
- Date of Event
- June 12, 2015
- Report Date
- June 29, 2015
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS
- Product Code
- KHP
- PMA / PMN Number
- K113216
- Removal / Correction Number
- 1217157-07-24-2015-003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SIEMENS ISSUED AN URGENT FIELD SAFETY NOTICE (UFSN) 33114 REV. A IN JULY 2014 TO NOTIFY ALL AFFECTED SIEMENS CUSTOMERS ABOUT THE ISSUE. THE SECTION 'ACTIONS TO BE TAKEN BY THE CUSTOMER' OF UFSN WILL INFORM CUSTOMERS TO DISABLE ONE PORT (EITHER THE ETHERNET PORT OR THE SERIAL PORT) ON THE RAPIDPOINT 500 ANALYZER. THE RESOLUTION OF THIS ISSUE WILL BE IMPLEMENTED IN THE NEXT SOFTWARE RELEASE FOR THE RAPID-POINT 500 SYSTEM.
AFTER SEVERAL EMAIL EXCHANGES WITH CUSTOMER, SIEMENS CONFIRMED THAT THE SERIAL PORT IS CONNECTED TO THE LIS AND IS THE SECONDARY PORT WHILE THE NETWORK PORT IS CONNECTED TO RAPIDCOMM AND IS THE PRIMARY PORT. THE ACTUAL CONNECTION TO THE LIS SYSTEM GOES THROUGH A DEVICE CALLED A 401P WHICH SPLITS THE LIS-3 SIGNAL AND SENDS A REFORMATTED LIS-2 MESSAGE TO THE LIS AND AN LIS-3 MESSAGE TO A TICKET PRINTER WHICH IS USED BY THEM TO OUTPUT SAMPLE RESULTS. THE CORRUPTED MESSAGE IS FROM THE RECEIVED LOGGING ON THE LIS SYSTEM. SINCE SOFTWARE V2.2, THE RP500 HAS DUAL PORT LIS TRANSMISSION CAPABILITY MEANING THE SYSTEM CAN BE CONNECTED TO 2 DATA MANAGEMENT SYSTEMS SUCH AS RAPIDCOMM AND ANOTHER DATA MANAGEMENT SYSTEM(DMS). THIS ALLOWS THE SYSTEM TO COMMUNICATE SIMULTANEOUSLY OVER THE SERIAL AND ETHERNET PORTS AND CONFIGURE EACH PORT TO SEND PATIENT, CALIBRATION, OR QC DATA. USER CAN SELECT WHICH DATA IS SENT OVER WHICH PORT AND CAN SPECIFY WHICH PORT SERVES AS THE PRIMARY INTERFACE. TYPICALLY A USER WOULD SETUP ONE PORT AS THE PRIMARY PORT WHICH WOULD INTERFACE WITH RCOMM OR A LIS AND SEND OUT ALL THE NECESSARY DATA. THEY WOULD THEN USE THE OTHER PORT TO JUST SEND OUT RESULTS, NOT HAVE IT SETUP TO ANY TYPE OF LIS. IN THIS COMPLAINT, THE CUSTOMER IS USING THIS DUAL TRANSMISSION FEATURE BUT BECAUSE THEY SETUP BOTH PORTS TO SEND OUT THE SAME TYPE OF INFORMATION, THE DATA IS GETTING DUPLICATED OR MISSED. SIEMENS INVESTIGATED THE ISSUE AND HAS IDENTIFIED THE ROOT CAUSE; MITIGATIONS WILL BE IMPLEMENTED IN NEW SOFTWARE VERSION.
CUSTOMER REPORTED THAT A PCO2 RESULT WAS NOT SENT TO LIS (DATA MANAGEMENT SYSTEM). CUSTOMER INDICATED THAT SYSTEM TRANSFERRED HCT RESULT TWICE AND DID NOT SEND PCO2 RESULT IN ONE PATIENT REPORT TO LIS. CUSTOMER ALSO INDICATED THAT THIS ISSUE OCCURRED ONLY ONE TIME SO FAR. THERE WAS NO REPORT OF SERIOUS INJURY DUE TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 487806 | RAPIDPOINT 500 | RP 500 | KHP | SIEMENS HEALTHCARE DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |