FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4942686 · Received July 27, 2015

Report

Report Number
3004209178-2015-77975
Event Type
Malfunction
Date Received
July 27, 2015
Date of Event
April 1, 2015
Report Date
July 6, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH INTERMITTENT ESC, ACT, UP ARROW, AND DOWN ARROW BUTTON RESPONSE DUE TO FLATTENED BUTTON DOME SWITCHES. INSULIN PUMP PASSED THE DISPLACEMENT, PRIME AND EXCESSIVE NO DELIVERY TEST. NO EXCESSIVE NO DELIVERY ALARM NOTED. INSULIN PUMP RECEIVED WITH MINOR SCRATCHED LCD WINDOW AND SCRATCHED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED VIA PHONE CALL, THAT THE BUTTONS ON THE INSULIN PUMP ARE UNRESPONSIVE. IT WAS ALSO REPORTED THAT THE CUSTOMER HAD A NO DELIVERY ALARM. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME WAS UNKNOWN. TROUBLESHOOTING OCCURRED. ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP, AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.