FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4942615
·
Received July 27, 2015
Report
- Report Number
- 3004209178-2015-77921
- Event Type
- Malfunction
- Date Received
- July 27, 2015
- Date of Event
- July 1, 2015
- Report Date
- July 6, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UNIT RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM DURING TESTING. UNIT RECEIVED WITH CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND CRACKED BELT CLIP SLOT.
Description of Event or Problem · 1
IT WAS REPORTED VIA PHONE CALL, THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 180 MG/DL. TROUBLESHOOTING OCCURRED. ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP, AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 487462 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |