FDA Adverse Event Malfunction Summary report: N

EB010, VOYANT 5MM FUSION, 37CM 6/BX

MDR report key: 4940833 · Received July 25, 2015

Report

Report Number
2027111-2015-00492
Event Type
Malfunction
Date Received
July 25, 2015
Date of Event
April 21, 2015
Report Date
July 24, 2015
Manufacturer
APPLIED MEDICAL
Product Code
GEI
PMA / PMN Number
K141288
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: THE EVENT UNIT WAS RETURNED FOR EVALUATION. ENGINEERING PERFORMED VISUAL AND FUNCTIONAL TESTING ON THE DEVICE. UPON INSPECTION, THE DEVICE MET ALL MECHANICAL SPECIFICATIONS BUT DID NOT MEET THE ELECTRICAL CONTINUITY REQUIREMENTS. DURING TESTING, ENGINEERING FOUND THAT THE DISTAL CONDUCTIVE STOP WAS SHORTING TO THE LOWER JAW. THIS WOULD RESULT IN "CHECK DEVICE" ALARMS DESCRIBED IN THE INCIDENT EXPERIENCE. ENGINEERING WAS ABLE TO REPLICATE THE ALARMS WHEN THE DEVICE WAS ACTIVATED AND ANALYSIS OF THE DEVICE KEY ACTIVATION LOGS ALSO CONFIRMED THE INCIDENT EXPERIENCE. THE ROOT CAUSE OF THE INCIDENT WAS DUE TO AN ELECTRICAL SHORT BETWEEN THE DISTAL CONDUCTIVE STOP AND THE LOWER JAW. APPLIED MEDICAL CONTINUOUSLY SEEKS TO IMPROVE THE FORM, FUNCTION, AND EASE OF USE OF ITS PRODUCTS. AS A RESULT OF THIS FEEDBACK, ADDITIONAL INSPECTION STEPS AND PROCESS ENHANCEMENTS HAVE BEEN ADDED TO THE MANUFACTURING PROCESS TO FURTHER MINIMIZE THE POTENTIAL FOR THIS TYPE OF INCIDENT TO REOCCUR. THIS DOCUMENT REPRESENTS OUR INITIAL/FINAL REPORT.

Description of Event or Problem · 1

T.L.H.( HYSTERECTOMIE) :" STARTED WELL, SEEMS TO BLEED A BIT MORE COMPARED TO OTHER DEVICES. THEN, WE'VE GOT 10 NO PROBLEM DURING THE TLH SUDDENLY, WE'VE GOT SEVERAL ALARMS, THE KNOW OF "CLEAN INSTRU; LITTLE BIT OF TISSUE...." THEN IT WORK FOR ANOTHER 10 SEALS VERY GOOD AND AGAIN SOME ALARM." PATIENT STATUS:" NO PROBLEMS."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
484908 EB010, VOYANT 5MM FUSION, 37CM 6/BX GEI GEI APPLIED MEDICAL EB010 1240553

Patients

Seq Age Sex Outcome Treatment
1 Other