FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 4940283 · Received July 24, 2015

Report

Report Number
2531779-2015-25473
Event Type
Malfunction
Date Received
July 24, 2015
Report Date
July 6, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 09/16/2015-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE PUMP POWERED ON TO THE VERIFY SCREEN WITH A FULLY ILLUMINATED AND LEGIBLE DISPLAY. THE COMPLAINT OF A BLANK SCREEN ISSUE WAS NOT DUPLICATED DURING EVALUATION. THERE WERE NO DEFECTS WERE FOUND. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN) ISSUE. THE DISTRIBUTOR ALLEGED THAT THE DISPLAY SCREEN WAS BLANK. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
483765 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1