FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 4939752
·
Received July 24, 2015
Report
- Report Number
- 3008642652-2015-04622
- Event Type
- Malfunction
- Date Received
- July 24, 2015
- Date of Event
- July 1, 2015
- Report Date
- July 24, 2015
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESETS) WAS CONFIRMED. UPON EVALUATION THE MONITOR WAS RESETTING DURING PULSE TESTING. THE ROOT CAUSE FOR THE RESETS WAS ISOLATED TO NOISE FROM THE DEFIBRILLATOR PCA HIGH-VOLTAGE CAPACITORS PROPAGATING ON THE MAIN BATTERY WIRE ON THE MONITOR C/A BOARD. PMA SUPPLEMENT P010030/S064 WAS SUBMITTED TO FDA ON 07/06/2015 AND IS CURRENTLY UNDER REVIEW. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE MONITOR.
Description of Event or Problem · 1
A US DISTRIBUTOR RETURNED MONITOR SN (B)(4) AND REPORTED THAT THE MONITOR WAS RESETTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 483382 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |