FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4939063
·
Received July 24, 2015
Report
- Report Number
- 3004209178-2015-77263
- Event Type
- Malfunction
- Date Received
- July 24, 2015
- Date of Event
- July 17, 2015
- Report Date
- July 17, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INSULIN PUMP RECEIVED WITH BUTTON ERROR ALARM AND HAD UNRESPONSIVE BUTTONS DUE TO CORRODED KEYPAD TRACES. UNABLE TO PERFORM OPERATING CURRENT TEST OR VERIFY BATTERY OUT LIMIT DUE TO UNRESPONSIVE BUTTONS. INSULIN PUMP HAD CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.
Description of Event or Problem · 1
CUSTOMER CALLED TO REPORT THAT THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER STATED THAT THE INSULIN PUMP MAY HAVE BEEN EXPOSED TO MOISTURE. CUSTOMER ALSO REPORTED THAT THE INSULIN PUMP ALARMED BATTERY OUT LIMIT. CUSTOMER'S BLOOD GLUCOSE READING WAS 85 MG/DL. ADVISED DISCONTINUATION OF THE INSULIN PUMP AND REVERT TO BACK UP PLAN PER HEALTH CARE PROFESSIONAL. PRODUCT IS BEING RETURNED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 484065 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |