FDA Adverse Event Injury Summary report: N

BONE PROBE CURVED LENKE

MDR report key: 4934555 · Received July 21, 2015

Report

Report Number
3005739886-2015-00058
Event Type
Injury
Date Received
July 21, 2015
Date of Event
June 23, 2015
Report Date
June 23, 2015
Manufacturer
PRECISION SPINE, INC.
Product Code
HXB
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION.

Additional Manufacturer Narrative · 1

ENGINEERING REVIEW NOTED THE PSSPCL CURVED LENKE BONE PROBE FRACTURED DURING THE INTENDED USE OF THIS INSTRUMENT. REVIEW OF THE COMPLAINT SUMMARY INDICATES THAT THE BONE PROBE WAS REPEATEDLY STRUCK WITH A MALLET TO ADVANCE THE PROBED THROUGH VERY HARD BONE. THE EXCESSIVE FORCE OF IMPACTION WAS THE PROBABLE CAUSE OF THE FAILURE OF THE TIP AND CAN BE ATTRIBUTED TO MISUSE. REVIEW OF MANUFACTURING HISTORY RECORDS FOUND A TOTAL OF FIVE (5) PIECES OF THIS LOT WERE RELEASED FOR DISTRIBUTION ON APRIL 30, 2008 WITH NO DEVIATION OR ANOMALIES. A TWO-YEAR COMPLAINT HISTORY REVIEW DID NOT IDENTIFY A TREND OF MISUSE FOR THIS PART NUMBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE PERFORMED (B)(6) 2015, ON A PATIENT WITH VERY HARD BONE, THE DOCTOR HAD TO TRIKE THE BONE PROBE REPEATEDLY WITH A MALLET, RESULTING IN BREAKAGE OF THE TIP OF THE PROBE. APPROXIMATELY 20MM OF THE TIP REMAINED IN THE PEDICLE. THE DOCTOR USED A TREPHINE TO BURR AROUND THE TIP IN ORDER TO GRAB AND PULL IT OUT. A DELAY OF APPROXIMATELY 40 MINUTES RESULTED WHILE A TOOL WAS LOCATED AND STERILIZED TO BE USED TO PULL THE TIP OUT OF THE BONE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE PERFORMED (B)(6) 2015 ON A PATIENT WITH VERY HARD BONE, THE DOCTOR HAD TO STRIKE THE BONE PROBE REPEATEDLY WITH A MALLET, RESULTING IN BREAKAGE OF THE TIP OF THE PROBE. APPROXIMATELY 20MM OF THE TIP REMAINED IN THE PEDICLE. THE DOCTOR USED A TREPHINE TO BURR AROUND THE TIP IN ORDER TO GRAB AND PULL IT OUT. A DELAY OF APPROXIMATELY 40 MINUTES RESULTED WHILE A TOOL WAS LOCATED AND STERILIZED TO BE USED TO PULL THE TIP OUT OF THE BONE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
473172 BONE PROBE CURVED LENKE HXB HXB PRECISION SPINE, INC. 590223D08

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention