BONE PROBE CURVED LENKE
Report
- Report Number
- 3005739886-2015-00058
- Event Type
- Injury
- Date Received
- July 21, 2015
- Date of Event
- June 23, 2015
- Report Date
- June 23, 2015
- Manufacturer
- PRECISION SPINE, INC.
- Product Code
- HXB
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION.
ENGINEERING REVIEW NOTED THE PSSPCL CURVED LENKE BONE PROBE FRACTURED DURING THE INTENDED USE OF THIS INSTRUMENT. REVIEW OF THE COMPLAINT SUMMARY INDICATES THAT THE BONE PROBE WAS REPEATEDLY STRUCK WITH A MALLET TO ADVANCE THE PROBED THROUGH VERY HARD BONE. THE EXCESSIVE FORCE OF IMPACTION WAS THE PROBABLE CAUSE OF THE FAILURE OF THE TIP AND CAN BE ATTRIBUTED TO MISUSE. REVIEW OF MANUFACTURING HISTORY RECORDS FOUND A TOTAL OF FIVE (5) PIECES OF THIS LOT WERE RELEASED FOR DISTRIBUTION ON APRIL 30, 2008 WITH NO DEVIATION OR ANOMALIES. A TWO-YEAR COMPLAINT HISTORY REVIEW DID NOT IDENTIFY A TREND OF MISUSE FOR THIS PART NUMBER.
IT WAS REPORTED THAT DURING A PROCEDURE PERFORMED (B)(6) 2015, ON A PATIENT WITH VERY HARD BONE, THE DOCTOR HAD TO TRIKE THE BONE PROBE REPEATEDLY WITH A MALLET, RESULTING IN BREAKAGE OF THE TIP OF THE PROBE. APPROXIMATELY 20MM OF THE TIP REMAINED IN THE PEDICLE. THE DOCTOR USED A TREPHINE TO BURR AROUND THE TIP IN ORDER TO GRAB AND PULL IT OUT. A DELAY OF APPROXIMATELY 40 MINUTES RESULTED WHILE A TOOL WAS LOCATED AND STERILIZED TO BE USED TO PULL THE TIP OUT OF THE BONE.
IT WAS REPORTED THAT DURING A PROCEDURE PERFORMED (B)(6) 2015 ON A PATIENT WITH VERY HARD BONE, THE DOCTOR HAD TO STRIKE THE BONE PROBE REPEATEDLY WITH A MALLET, RESULTING IN BREAKAGE OF THE TIP OF THE PROBE. APPROXIMATELY 20MM OF THE TIP REMAINED IN THE PEDICLE. THE DOCTOR USED A TREPHINE TO BURR AROUND THE TIP IN ORDER TO GRAB AND PULL IT OUT. A DELAY OF APPROXIMATELY 40 MINUTES RESULTED WHILE A TOOL WAS LOCATED AND STERILIZED TO BE USED TO PULL THE TIP OUT OF THE BONE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 473172 | BONE PROBE CURVED LENKE | HXB | HXB | PRECISION SPINE, INC. | 590223D08 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |