FDA Adverse Event Injury Summary report: N

PUMP IN STYLE ADVANCED BREAT PUMP STARTER

MDR report key: 4932419 · Received July 22, 2015

Report

Report Number
1419937-2015-00231
Event Type
Injury
Date Received
July 22, 2015
Date of Event
July 7, 2015
Report Date
July 22, 2015
Manufacturer
MEDELA INC.
Product Code
HGX
PMA / PMN Number
K031614
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT PUMP WAS SENT TO THE CUSTOMER. A MEDELA CLINICIAN FOLLOWED UP WITH THE CUSTOMER ON 07/09/2015. THE CUSTOMER STATED THAT SHE DEVELOPED MASTITIS AND WAS PRESCRIBED DICLOXACILLIN BY HER PHYSICIAN, WHILE USING THE PUMP IN STYLE. IT CANNOT BE DEFINITIVELY CONCLUDED THAT THE PUMP CAUSED OR CONTRIBUTED TO THE CUSTOMER'S MASTITIS. REPORTED ISSUES OF MASTITIS ARE UNDER INVESTIGATION IN (B)(4). MASTITIS IS USUALLY A BENIGN, SELF-LIMITING INFECTION WITH FEW CONSEQUENCES FOR THE SUCKLING INFANT. THE RISK OF MASTITIS IS HIGHER AMONG WOMEN WHO HAVE BREASTFED PREVIOUSLY, ESPECIALLY THOSE WITH A HISTORY OR MASTITIS." RIORDAN & WAMBACH, 4TH ED. P. 294: BREASTFEEDING AND HUMAN LACTATION. MASTITIS REQUIRES PROMPT MEDICAL ATTENTION FOR THE MOTHER FOR PAIN RELIEF AND PRESCRIPTION ANTIBIOTICS TO AVOID PROGRESSION TO OVERWHELMING SEPSIS. THE PRODUCT INVOLVED IN THE COMPLAINT WAS NOT RETURNED FOR EVALUATION/INVESTIGATION AT THIS TIME. THEREFORE, NO CONCLUSIONS CAN BE MADE AS TO THE CAUSE OF THE EVENT. SHOULD ADDITIONAL INFORMATION OR THE ORIGINAL PRODUCT BE RECEIVED, RESULTING IN NEW, CHANGED, OR CORRECTED INFORMATION, A FOLLOW UP REPORT WILL BE FILED AT THAT TIME.

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO MEDELA AND EVALUATED ON 10/08/2015; THE EVALUATION SHOWED THAT THE PUMP PASSED ALL FUNCTIONAL TESTS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO CUSTOMER SERVICE THAT HER PUMP IN STYLE BREAST PUMP WAS MAKING A KNOCKING NOISE AND HAS LOW SUCTION. SHE ALSO STATED THAT SHE DEVELOPED MASTITIS DURING THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
476001 PUMP IN STYLE ADVANCED BREAT PUMP STARTER HGX HGX MEDELA INC. 57081 415067

Patients

Seq Age Sex Outcome Treatment
1 Other