SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Report
- Report Number
- 1416980-2015-29754
- Event Type
- Malfunction
- Date Received
- July 22, 2015
- Date of Event
- July 1, 2015
- Report Date
- July 1, 2015
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS REPORTED. THE DEVICE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR 2240 ALARM (AIR IN SET/LINE) OCCURRED ON THE HOMECHOICE DEVICE DURING DWELL FOUR OF FOUR IN PERITONEAL DIALYSIS. THE HOME PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THE TECHNICAL SERVICE REPRESENTATIVE EXPLAINED THE ALARM TO THE HOME PATIENT. THE TECHNICAL SERVICE REPRESENTATIVE HAD THE HOME PATIENT CYCLE THE POWER TO CLEAR THE ALARM AND ASSISTED TO RETRIEVE THE CASSETTE. THERE WAS NOTHING UNUSUAL NOTED WITH THE SUPPLIES OR THE SETUP THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE ALARM. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 476213 | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | HOMECHOICE |