FDA Adverse Event Malfunction Summary report: N

MOTOR DRIVE UNIT,ULTRALIGHT W/

MDR report key: 4931842 · Received July 22, 2015

Report

Report Number
1643264-2015-00044
Event Type
Malfunction
Date Received
July 22, 2015
Date of Event
April 19, 2013
Report Date
July 22, 2015
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HAB
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION OF THE RETURNED DEVICE CONFIRMED THE REPORTED COMPLAINT. THE MOTOR ((B)(4)) STALLS AT LESS THAN 500 RPM. MOTOR HAS FAILED UNDER WARRANTY AND WILL BE RETURNED TO THE SUPPLIER. SMITH & NEPHEW WILL CONTINUE TO MONITOR FOR ANY FUTURE OCCURENCE OF THIS FAILURE TYPE. (B)(4).

Description of Event or Problem · 1

DURING AN UNKNOWN PROCEDURE THE MOTOR DRIVE UNIT, ULTRALIGHT W/ ELECTRICAL GETS HEATED WHEN USED IN SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
477059 MOTOR DRIVE UNIT,ULTRALIGHT W/ SAW, POWERED, AND ACCESSORIES HAB SMITH & NEPHEW, INC.

Patients

Seq Age Sex Outcome Treatment
1