FDA Adverse Event
Malfunction
Summary report: N
MOTOR DRIVE UNIT,ULTRALIGHT W/
MDR report key: 4931842
·
Received July 22, 2015
Report
- Report Number
- 1643264-2015-00044
- Event Type
- Malfunction
- Date Received
- July 22, 2015
- Date of Event
- April 19, 2013
- Report Date
- July 22, 2015
- Manufacturer
- SMITH & NEPHEW, INC.
- Product Code
- HAB
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION OF THE RETURNED DEVICE CONFIRMED THE REPORTED COMPLAINT. THE MOTOR ((B)(4)) STALLS AT LESS THAN 500 RPM. MOTOR HAS FAILED UNDER WARRANTY AND WILL BE RETURNED TO THE SUPPLIER. SMITH & NEPHEW WILL CONTINUE TO MONITOR FOR ANY FUTURE OCCURENCE OF THIS FAILURE TYPE. (B)(4).
Description of Event or Problem · 1
DURING AN UNKNOWN PROCEDURE THE MOTOR DRIVE UNIT, ULTRALIGHT W/ ELECTRICAL GETS HEATED WHEN USED IN SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 477059 | MOTOR DRIVE UNIT,ULTRALIGHT W/ | SAW, POWERED, AND ACCESSORIES | HAB | SMITH & NEPHEW, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|