DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2015-00167
- Event Type
- Malfunction
- Date Received
- July 21, 2015
- Date of Event
- July 13, 2015
- Report Date
- September 17, 2015
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- JGJ
- PMA / PMN Number
- K140790
- Removal / Correction Number
- 2517506-09292015-006-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ORIGINAL MDR SUBMITTED (B)(6) 2015. SIEMENS HEALTHCARE DIAGNOSTICS HAS RECEIVED CUSTOMER COMPLAINTS FOR THE DIMENSION VISTA MAGNESIUM (MG) LOT 15063BA. THE COMPLAINTS WERE RELATED TO ERRONEOUS LOW RESULTS ON A SMALL NUMBER OF REAGENT WELLS. SIEMENS HEALTHCARE DIAGNOSTICS ISSUED AN URGENT MEDICAL DEVICE RECALL COMMUNICATION VC-15-12.A.US DATED SEPTEMBER 2015 INSTRUCTING CUSTOMERS TO DISCONTINUE USE AND DISCARD ANY REMAINING INVENTORY OF THE AFFECTED LOT AND TO CONTACT SIEMENS HEALTHCARE DIAGNOSTICS FOR REPLACEMENT PRODUCT. DATA COLLECTED FROM THE CUSTOMER COMPLAINTS SHOWS THAT THERE IS THE POTENTIAL FOR UNDER-RECOVERY ON CERTAIN WELLS FOR QUALITY CONTROL(QC) AND PATIENT SAMPLE RESULTS. WHEN THIS ISSUE OCCURRED, THE UNDER-RECOVERY RANGED FROM -0.3 MG/DL [0.12 MMOL/L] TO - 1.6 MG/DL [0.66 MMOL/L]. THE AVERAGE UNDER-RECOVERY WAS -1.0 MG/DL [0.41 MMOL/L]. THIS ISSUE IS DEMONSTRATING A LOW FREQUENCY OF OCCURRENCE WITH AN ESTIMATE ONE (1) IN EVERY 10,000 TESTS MAY BE AFFECTED. REPEATING THE SAMPLE ON A NEW WELL OR A NEW FLEX CARTRIDGE RESOLVES THE ISSUE. THIS ISSUE IS LIKELY TO BE DETECTED BY QUALITY CONTROL (QC) OR CALIBRATORS IF THEY ARE RUN ON THE AFFECTED WELLS OF REAGENT.
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED MG RESULTS IS UNKNOWN. HOWEVER, THE INSTRUMENT DATA SHOWS THE CLUSTER OF DEPRESSED MG RESULTS WAS RESTRICTED TO A REAGENT WELL PAIR IN THE MG FLEX REAGENT CARTRIDGE. THE ISSUE IS UNDER INVESTIGATION BY (B)(4).
DISCREPANT DEPRESSED MAGNESIUM (MG) RESULTS WERE OBTAINED ON PATIENT SAMPLES. IT IS UNKNOWN IF PATIENT RESULTS WERE REPORTED TO PHYSICIANS. THE ACCOUNT REPEATED THE SAMPLES ON AN ALTERNATE REAGENT WELL SET OF THE SAME FLEX(R) REAGENT CARTRIDGE AND HIGHER RESULTS WERE OBTAINED. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PRESCRIBED ON THE BASIS OF THE FALSELY DEPRESSED MG RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED MG RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 474098 | DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM | MAGNESIUM FLEX® REAGENT CARTRIDGE | JGJ | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | 15063BA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |