FDA Adverse Event Injury Summary report: N

VASCULAR POSITIONING SYSTEM

MDR report key: 4929854 · Received July 21, 2015

Report

Report Number
2518433-2015-00007
Event Type
Injury
Date Received
July 21, 2015
Date of Event
July 7, 2015
Report Date
July 7, 2015
Manufacturer
VASONOVA INC.
Product Code
OBJ
PMA / PMN Number
K123813
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED COMPLAINT WAS CONFIRMED THROUGH EVALUATION OF THE RETURNED CONSOLE AND DATASET. THE X-RAY MENTIONED IN THE INITIAL REPORT WAS NOT PROVIDED. THE CONSOLE WAS EXAMINED AND FOUND TO BE IN ACCEPTABLE CONDITION. THE UNIT WAS TESTED AND FUNCTIONED AS INTENDED AND A REVIEW OF MANUFACTURING RECORDS DID NOT YIELD ANY RELEVANT FINDINGS. THE SENIOR ALGORITHM SCIENTIST REVIEWED THE SAVED PROCEDURE DATASET AND DETERMINED THAT THE OPERATOR CONTINUED TO ADVANCE THE STYLET IN THE DIRECTION OF THE HEART EVEN THOUGH THE "ORANGE STOP" SYMBOL AND ARTERIAL FLOW DOPPLER WERE CONTINUOUSLY DISPLAYED AFTER INITIAL PLACEMENT IN THE PERIPHERAL SITE. THIS WAS CONTRARY TO THE INSTRUCTIONS IN THE PROVIDED OPERATOR'S MANUAL . THEREFORE, USE ERROR CAUSED OR CONTRIBUTED TO THIS EVENT. A CUSTOMER IN-SERVICE WAS REQUESTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS BEING PERFORMED IN THE INTENSIVE CARE UNIT. DURING INSERTION OF A BARD 5FR TRIPLE LUMEN CATHETER THE CLINICIAN RECEIVED A BLUE BULLSEYE WITH THE VPS. ADDITIONAL INFORMATION PROVIDED BY THE TELEFLEX CLINICAL EDUCATION MANAGER INDICATES THE CLINICIAN TURNED ON VPS UNIT WHEN THEY WERE 10CM INTO THE VESSEL. INITIAL SYMBOL WAS GREEN WITH LOW AMPLITUDE ANTE-GRADE DOPPLER, THEN IT WENT ORANGE WITH RETROGRADE FLOW. KEPT TOGGLING BETWEEN GREEN AND ORANGE SYMBOLS. AFTER SEVERAL FRAMES OF RETROGRADE FLOW WITH ORANGE SYMBOL, SYMBOL TURNED GREEN AND DOPPLER BECAME HIGH AMPLITUDE ANTE-GRADE FLOW FOR SEVERAL FRAMES. THERE WERE A COUPLE FLEETING BLUE BULLSEYES FOLLOWED BY A STEADY BLUE BULLSEYE FOR AT LEAST 10 SECONDS. OVER THE WEEKEND, A SUBSEQUENT CHEST X-RAY WAS TAKEN FOR ANOTHER REASON AND READING RADIOLOGIST NOTICED THAT THE PICCC WAS OVER THE CLAVICLE AND TERMINATED ON THE LEFT SIDE OF THE HEART INDICATING AORTA PLACEMENT. VASCULAR SURGERY WAS CONSULTED, PATIENT WAS FOUND WITH NO BRACHIAL PULSE. THE PATIENT WAS TAKEN TO SURGERY FOR A THROMBECTOMY AND CATHETER WAS REMOVED. THE PATIENT HAD NO FURTHER ADVERSE EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472425 VASCULAR POSITIONING SYSTEM VPS G4 SYSTEM OBJ VASONOVA INC.

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention