FDA Adverse Event Malfunction Summary report: N

DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM

MDR report key: 4929830 · Received July 21, 2015

Report

Report Number
2517506-2015-00161
Event Type
Malfunction
Date Received
July 21, 2015
Date of Event
June 23, 2015
Report Date
September 17, 2015
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Product Code
JGJ
PMA / PMN Number
K140790
Removal / Correction Number
2517506-09292015-006-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ORIGINAL MDR SUBMITTED 2015-07-21. SIEMENS HEALTHCARE DIAGNOSTICS HAS RECEIVED CUSTOMER COMPLAINTS FOR THE DIMENSION VISTA MAGNESIUM (MG) LOT 15063BA. THE COMPLAINTS WERE RELATED TO ERRONEOUS LOW RESULTS ON A SMALL NUMBER OF REAGENT WELLS. SIEMENS HEALTHCARE DIAGNOSTICS ISSUED AN URGENT MEDICAL DEVICE RECALL COMMUNICATION VC-15-12.A. US DATED SEPTEMBER 2015 INSTRUCTING CUSTOMERS TO DISCONTINUE USE AND DISCARD ANY REMAINING INVENTORY OF THE AFFECTED LOT AND TO CONTACT SIEMENS HEALTHCARE DIAGNOSTICS FOR REPLACEMENT PRODUCT. DATA COLLECTED FROM THE CUSTOMER COMPLAINTS SHOWS THAT THERE IS THE POTENTIAL FOR UNDER-RECOVERY ON CERTAIN WELLS FOR QUALITY CONTROL(QC) AND PATIENT SAMPLE RESULTS. WHEN THIS ISSUE OCCURRED, THE UNDER-RECOVERY RANGED FROM -0.3 MG/DL [0.12 MMOL/L] TO - 1.6 MG/DL [0.66 MMOL/L]. THE AVERAGE UNDER-RECOVERY WAS -1.0 MG/DL [0.41 MMOL/L]. THIS ISSUE IS DEMONSTRATING A LOW FREQUENCY OF OCCURRENCE WITH AN ESTIMATE ONE (1) IN EVERY 10,000 TESTS MAY BE AFFECTED. REPEATING THE SAMPLE ON A NEW WELL OR A NEW FLEX CARTRIDGE RESOLVES THE ISSUE. THIS ISSUE IS LIKELY TO BE DETECTED BY QUALITY CONTROL (QC) OR CALIBRATORS IF THEY ARE RUN ON THE AFFECTED WELLS OF REAGENT.

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED MG RESULTS IS UNKNOWN. HOWEVER, THE INSTRUMENT DATA SHOWS THE CLUSTER OF DEPRESSED MG RESULTS WAS RESTRICTED TO A REAGENT WELL PAIR IN THE MG FLEX REAGENT CARTRIDGE. THE ISSUE IS UNDER INVESTIGATION BY SIEMENS HEALTHCARE DIAGNOSTICS INC..

Description of Event or Problem · 1

DISCREPANT DEPRESSED MAGNESIUM (MG) RESULTS WERE OBTAINED ON PATIENT SAMPLES. PATIENT RESULTS WERE NOT REPORTED TO PHYSICIANS. THE ACCOUNT REPEATED THE SAMPLES ON AN ALTERNATE DIMENSION VISTA INSTRUMENT WITH A DIFFERENT MG REAGENT LOT AND HIGHER RESULTS WERE OBTAINED AND REPORTED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED ON THE BASIS OF THE FALSELY DEPRESSED MG RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED MG RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
473352 DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM MAGNESIUM FLEX® REAGENT CARTRIDGE JGJ SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW 15063BA

Patients

Seq Age Sex Outcome Treatment
1