FDA Adverse Event Malfunction Summary report: N

TRANSURETHRAL CATHETER

MDR report key: 4929649 · Received July 21, 2015

Report

Report Number
8040412-2015-00157
Event Type
Malfunction
Date Received
July 21, 2015
Date of Event
June 1, 2015
Report Date
June 25, 2015
Manufacturer
TELEFLEX MEDICAL
Product Code
FGH
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

QN#(B)(4). THE DEVICE SAMPLE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR INVESTIGATION AT THE TIME OF THIS REPORT. THE MANUFACTURER WILL CONTINUE TO MONITOR AND TREND RELATED EVENTS.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HISTORY RECORD FOR LOT 14KE45 WAS REVIEWED AND NO ISSUE THAT COULD HAVE CONTRIBUTED TO THE REPORTED FAILURE WAS NOTED. THE DEVICE WAS MANUFACTURED ACCORDING TO RELEASE SPECIFICATION. THE RETURNED SAMPLE WAS 100% VISUALLY INSPECTED USING MAGNIFICATION LENS. THE SAMPLE APPEARED TO BE IN A GOOD CONDITION EXCEPT THAT THE BALLOON WAS SPLIT. CLOSER EXAMINATION ON THE SPLIT BALLOON AREA UNDER DINO-LITE REVEALED SCRATCH MARKS NEAR THE TEAR REGION. ANY DEFECTIVE RAW BALLOON WILL BE CULLED OUT BEFORE BEING SENT TO THE NEXT PROCESS. AFTER ASSEMBLY, ALL FOLEY CATHETERS ARE THEN SUBJECTED TO LEAK TEST. PRODUCTS THAT PASS THIS TEST WILL BE SUBJECTED TO THE NEXT PROCESS. BASED ON THE INVESTIGATION CONDUCTED ON THE ACTUAL SAMPLE, IT IS BELIEVED THAT THE SCRATCH MARKS FOUND NEAR THE SPLIT AREA HAD INITIATED THE TEAR, DUE TO CONTACT WITH SHARP OR POINTED OBJECT. THEREFORE, WE COULD NOT CONFIRM THIS COMPLAINT, AS STATED.

Description of Event or Problem · 1

ALLEGED EVENT: THE NURSE INSERTED THE CATHETER, WHEN BALLOON WAS INFLATED, IT SNAPPED. THE PATIENT'S CONDITION WAS REPORTED AS FINE.

Description of Event or Problem · 1

ALLEGED EVENT: THE NURSE INSERTED THE CATHETER, WHEN BALLOON WAS INFLATED, IT SNAPPED. THE PATIENT'S CONDITION WAS REPORTED AS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470849 TRANSURETHRAL CATHETER FOLEY CATHETER FGH TELEFLEX MEDICAL 14KE45

Patients

Seq Age Sex Outcome Treatment
1