FDA Adverse Event Malfunction Summary report: N

VASCULAR PROBES

MDR report key: 4928319 · Received July 21, 2015

Report

Report Number
1416980-2015-29551
Event Type
Malfunction
Date Received
July 21, 2015
Report Date
September 10, 2015
Manufacturer
SYNOVIS SURGICAL INNOVATIONS
Product Code
DWP
PMA / PMN Number
K130896
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED AND REVEALED A BROWN FIBER APPROXIMATELY 1.8 MILLIMETERS LONG WITHIN THE INNER POUCH NEAR THE DEVICE TIP. THE PHYSICAL PROPERTIES OF THE FIBER ARE CONSISTENT WITH A CORRUGATED CARDBOARD PARTICLE. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION COULD NOT BE DETERMINED. A CAPA HAS BEEN OPENED TO ADDRESS THIS ISSUE. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PARTICULATE MATTER WAS FOUND IN THE INNER POUCH OF A VASCULAR PROBE. THIS WAS IDENTIFIED DURING RECEIVING INSPECTION BY THE REPORTING FACILITY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471699 VASCULAR PROBES DILATOR, VESSEL, SURGICAL DWP SYNOVIS SURGICAL INNOVATIONS SP15D091043706