VASCULAR PROBES
Report
- Report Number
- 1416980-2015-29551
- Event Type
- Malfunction
- Date Received
- July 21, 2015
- Report Date
- September 10, 2015
- Manufacturer
- SYNOVIS SURGICAL INNOVATIONS
- Product Code
- DWP
- PMA / PMN Number
- K130896
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED AND REVEALED A BROWN FIBER APPROXIMATELY 1.8 MILLIMETERS LONG WITHIN THE INNER POUCH NEAR THE DEVICE TIP. THE PHYSICAL PROPERTIES OF THE FIBER ARE CONSISTENT WITH A CORRUGATED CARDBOARD PARTICLE. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION COULD NOT BE DETERMINED. A CAPA HAS BEEN OPENED TO ADDRESS THIS ISSUE. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT PARTICULATE MATTER WAS FOUND IN THE INNER POUCH OF A VASCULAR PROBE. THIS WAS IDENTIFIED DURING RECEIVING INSPECTION BY THE REPORTING FACILITY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 471699 | VASCULAR PROBES | DILATOR, VESSEL, SURGICAL | DWP | SYNOVIS SURGICAL INNOVATIONS | SP15D091043706 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |