FDA Adverse Event Malfunction Summary report: N

MAQUET HEATER COOLER UNIT HCU30

MDR report key: 4927471 · Received July 21, 2015

Report

Report Number
8010762-2015-00579
Event Type
Malfunction
Date Received
July 21, 2015
Date of Event
February 26, 2013
Report Date
March 5, 2013
Manufacturer
MAQUET CARDIOPULMONARY AG
Product Code
DWC
PMA / PMN Number
K031544
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

"MAQUET MEDICAL SYSTEM, USA SUBMITS THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET CARDIOPULMONARY (B)(4). THE DEVICE WAS EVALUATED BY THE SSU AND THE COMPRESSOR WAS FOUND TO BE DEFECTIVE AND WAS REPLACED. (B)(4)."

Description of Event or Problem · 1

"ON (B)(6) 2013 THE SSU REPRESENTATIVE FOR MAQUET IN (B)(6) REPORTED THAT WITH CARDIOPLEGIA SIDE OF THE HCU 30 SERIAL NUMBER 100155 AT 8ºC AND PATIENT SIDE AT 38 ºC, THE TANK TEMPERATURE WAS AT 3ºC. AFTER 15 TO 30 MIN COMPRESSOR GOT TURNED OFF REGARDLESS THE TANK TEMPERATURE. THE TANK TEMPERATURE BEGAN TO RISE AS DID THE CARDIOPLEGIA SIDE WATER TEMPERATURE AND COULD NOT BE REGULATED. THERE WAS NO CHANGE TO THE PATIENT SIDE TEMPERATURE AND IT CONTINUED TO BE REGULATED. (B)(4)."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472442 MAQUET HEATER COOLER UNIT HCU30 CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC MAQUET CARDIOPULMONARY AG UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1