FDA Adverse Event Injury Summary report: N

EDWARDS EXPANDABLE INTRODUCER SHEATH SET

MDR report key: 4926561 · Received July 20, 2015

Report

Report Number
2015691-2015-01773
Event Type
Injury
Date Received
July 20, 2015
Date of Event
June 25, 2015
Report Date
June 25, 2015
Manufacturer
EDWARDS LIFESCIENCES
Product Code
NPT
PMA / PMN Number
P130009
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), CARDIOVASCULAR COMPLICATIONS, INCLUDING PERFORATION OR DISSECTION OF VESSELS WHICH MAY REQUIRE INTERVENTION, ARE POTENTIAL ADVERSE EVENTS ASSOCIATED WITH THE TRANSFEMORAL TRANSCATHETER AORTIC VALVE REPLACEMENT PROCEDURE. ACCORDING TO LITERATURE REVIEW, AND AS DOCUMENTED IN A TECHNICAL SUMMARY WRITTEN BY EDWARDS LIFESCIENCES, VASCULAR COMPLICATIONS ARE A WELL-RECOGNIZED COMPLICATION OF THE TRANSFEMORAL TAVR PROCEDURE IN THIS ELDERLY POPULATION WITH MULTIPLE CO-MORBIDITIES. EDWARDS HAS REVIEWED MANY REPORTS, INCLUDING SCREENING DATA RECORDS AND SOURCE DOCUMENTATION OF VASCULAR COMPLICATIONS AND HAS FOUND THAT THE ROOT CAUSE IS TYPICALLY RELATED TO A COMBINATION OF VESSEL SIZE, TORTUOSITY AND CALCIFICATIONS. ALTHOUGH THE INCIDENCE IS DECREASING WITH SMALLER SHEATH/DELIVERY SYSTEM SIZES AND PHYSICIAN EXPERIENCE, THERE WILL CONTINUE TO BE CASES IN WHICH VASCULAR COMPLICATIONS WILL OCCUR. THE THV PHYSICIAN TRAINING MANUALS INSTRUCT ON PROCEDURAL CONSIDERATIONS FOR SHEATH INSERTION WITH REGARDS TO PROPER SCREENING CRITICAL TO REDUCING VASCULAR COMPLICATIONS. THE TRAINING MANUAL INSTRUCTS THE OPERATOR ON PROPER SHEATH INSERTION AND WITHDRAWAL TECHNIQUES, INCLUDING PRE-DILATING THE VESSEL WITH THE EDWARDS DILATORS, AS NEEDED. IT ALSO NOTES THAT CALCIFICATION MAY REDUCE LUMEN DIAMETER AND LIMIT OR PREVENT TRANSFEMORAL PASSAGE OF THE DEVICES. THE IFU CONTRAINDICATES PATIENTS WITH ILIO-FEMORAL VESSEL CHARACTERISTICS THAT WOULD PRECLUDE SAFE PLACEMENT OF SHEATHS SUCH AS SEVERE OBSTRUCTIVE CALCIFICATION OR SEVERE TORTUOSITY. PRE-PROCEDURE SCREENING AND ASSESSMENT OF THE FEMORAL/ILIAC ARTERY INTERNAL DIAMETERS WILL ENABLE THE CLINICIAN TO DETERMINE IF THE SAPIEN XT VALVE CAN BE DELIVERED TRANSFEMORALLY. ASSESSMENT OF LOCATION AND AMOUNT OF CIRCUMFERENTIAL CALCIUM WILL AID IN DETERMINING AREAS OF REDUCED VESSEL DIAMETERS. THE OPERATORS ARE TRAINED TO MEASURE MINIMUM VESSEL DIAMETER TAKING CALCIUM INTO ACCOUNT. THE PHYSICIAN TRAINING MANUAL ALSO LISTS THE MINIMUM RECOMMENDED VESSEL SIZE FOR EACH SIZE DEVICE. DESPITE THE BEST SCREENING TOOLS, A SMALL PERCENTAGE OF PATIENTS WILL HAVE FEMORAL/ILIAC VESSELS THAT ARE NOT AMENABLE TO THE TRANS-FEMORAL APPROACH OR WHERE INCREASED RESISTANCE IS ENCOUNTERED DURING INSERTION OF DEVICES. IN MANY CASES, THE VESSEL MINIMUM LUMINAL DIAMETER (MLD) MAY BE BORDERLINE OR BELOW THE INDICATED SIZE. IN ADDITION, SIGNIFICANT CALCIFICATION AND/OR TORTUOSITY, NOT ALWAYS APPRECIABLE ON IMAGING, COULD BE CONTRIBUTING FACTORS TO THE EVENT. THE FOLLOWING GUIDANCE IS PROVIDED FOR SHEATH REMOVAL: ¿REMOVE THE ESHEATH ENTIRELY WITHOUT TORQUING ENSURING THE EDWARDS LOGO IS FACING UPWARDS¿. THE MINIMUM REQUIRED VESSEL DIAMETER FOR A 16FR ESHEATH IS 6.0MM. IN THIS CASE, THE CAUSE OF THE REPORTED ILIAC ARTERY DISSECTION CANNOT BE DETERMINED. PER REPORT, THE PATIENT ACCESS VESSEL MLD MEASURED 6.7MM WITH MILD CALCIFICATION AND TORTUOSITY. IT IS POSSIBLE THAT THERE MAY HAVE BEEN SOME VESSEL CALCIFICATION AND/OR TORTUOSITY NOT APPRECIABLE ON IMAGING THAT COULD HAVE CONTRIBUTED TO THE EVENT. THERE WAS NO ALLEGATION OR INDICATION THAT THE EVENT WAS RELATED TO A MANUFACTURING NON-CONFORMANCE. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED BY OUR (B)(4) AFFILIATE THAT DURING A TRANSFEMORAL TAVR PROCEDURE, WHEN WITHDRAWING THE ESHEATH, DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) SHOWED A DISSECTION BETWEEN THE RIGHT COMMON ILIAC ARTERY AND THE EXTERNAL ILIAC ARTERY. A 16FR DRY SHEATH WAS INSERTED AND A STENT GRAFT WAS PLACED. AFTER THE BLOOD FLOW WAS CONFIRMED, THE PROCEDURE WAS FINISHED. THE ACCESS VESSEL MINIMUM LUMINAL DIAMETER (MLD) MEASURED 6.7 MM WITH MILD CALCIFICATION AND TORTUOSITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469546 EDWARDS EXPANDABLE INTRODUCER SHEATH SET AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED NPT EDWARDS LIFESCIENCES 916ES23J 60054687

Patients

Seq Age Sex Outcome Treatment
1 94 YR Required Intervention