FDA Adverse Event Malfunction Summary report: N

ALPHATEC TARGETING NEEDLE, TROCAR TIP (STAINLESS STEEL)

MDR report key: 4925372 · Received July 20, 2015

Report

Report Number
2027467-2015-00153
Event Type
Malfunction
Date Received
July 20, 2015
Report Date
June 12, 2015
Manufacturer
ALPHATEC SPINE INC
Product Code
GAA
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MDR BEING SUBMITTED AS A RESULT OF A RETROSPECTIVE COMPLAINT REVIEW. DATE OF EVENT - HAS NOT BEEN PROVIDED AND IS UNKNOWN. LOT NUMBER - HAS NOT BEEN PROVIDED AND IS UNKNOWN. ALTHOUGH THE SUSPECT DEVICE HAS NOT BEEN RETURNED FOR EVALUATION, PERFORMANCE TESTING WAS CONDUCTED BY THE SUPPLIER OF ALPHATEC TARGETING NEEDLES. TESTING CONSISTED OF ULTIMATE PUSH (INSERTION OF THE DEVICE), ULTIMATE PULL (EXTRACTION OF THE DEVICE IF STUCK), AND ULTIMATE TORSION (TWISTING OF THE DEVICE FOR INSERTION OR EXTRACTION). THE SUPPLIER FOUND THAT DEVICE FUNCTIONS AS INTENDED. NO DISCREPANCIES OR ASSIGNABLE CAUSES WERE DETECTED DURING THEIR ROOT CAUSE INVESTIGATION.

Description of Event or Problem · 1

CHALLENGES THE FIELD HAS BEEN EXPERIENCED WITH TARGETING NEEDLES; 1) THE INNER STYLET GETS STUCK IN THE PEDICLE. 2) WHEN THE USER ATTEMPTS TO REMOVE THE OUTER CANNULA WITH THE T-HANDLE, THE HANDLE WILL COME OFF OF THE SHAFT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470220 ALPHATEC TARGETING NEEDLE, TROCAR TIP (STAINLESS STEEL) GAA GAA ALPHATEC SPINE INC 79701

Patients

Seq Age Sex Outcome Treatment
1 Other