FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 4924480 · Received July 16, 2015

Report

Report Number
1052693-2015-01140
Event Type
Malfunction
Date Received
July 16, 2015
Date of Event
June 17, 2015
Report Date
December 11, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT UNDER EVALUATION.

Additional Manufacturer Narrative · 1

(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USER'S TEST STRIP HAD POOR STORAGE. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF HIGH BLOOD GLUCOSE RESULTS. EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 120 TO 130 MG/DL. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THE TIME OF THE CALL ON (B)(6) 2015. VERIFIED STORAGE OF PRODUCT IS NOT WITHIN INSTRUCTED SPECIFICATION SINCE THEY ARE KEPT IN VEHICLE AND KITCHEN. TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 03/24/2018 AND OPEN VIAL DATE IS WEEK OF (B)(6) 2015. RECALL TEST RESULTS PERFORMED FASTING FROM METER MEMORY: 1:306MG/DL (B)(6) 2015 12:30PM. 2:437MG/DL (B)(6) 2015 11:43PM. 3:175MG/DL (B)(6) 2015 07:43AM . 4:160MG/DL (B)(6) 2015 06:26PM . 5:146MG/DL (B)(6) 2015 12:28PM . 6:192MG/DL (B)(6) 2015 07:52PM . 7:176MG/DL (B)(6) 2015 06:43PM. ADVERSE EVENT NOT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF HIGH BLOOD GLUCOSE RESULTS. EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 120 TO 130 MG/DL. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THE TIME OF THE CALL ON (B)(6) 2015. VERIFIED STORAGE OF PRODUCT IS NOT WITHIN INSTRUCTED SPECIFICATION SINCE THEY ARE KEPT IN VEHICLE AND KITCHEN. TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 03/24/2018 AND OPEN VIAL DATE IS WEEK OF (B)(6) 2015. RECALL TEST RESULTS PERFORMED FASTING FROM METER MEMORY: 306MG/DL, (B)(6) 2015, 12:30PM; 437MG/DL, (B)(6) 2015, 11:43PM; 175MG/DL, (B)(6) 2015, 07:43AM; 160MG/DL, (B)(6) 2015, 06:26PM; 146MG/DL, (B)(6) 2015, 12:28PM; 192MG/DL, (B)(6) 2015, 07:52PM; 176MG/DL, (B)(6) 2015, 06:43PM. ADVERSE EVENT NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463936 TRUERESULT BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PS2256

Patients

Seq Age Sex Outcome Treatment
1 0 YR