FDA Adverse Event Malfunction Summary report: N

TRUE RESULT

MDR report key: 4924475 · Received July 16, 2015

Report

Report Number
1052693-2015-01169
Event Type
Malfunction
Date Received
July 16, 2015
Date of Event
June 20, 2015
Report Date
July 16, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
LFR
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VT, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE: USER HAD AN INACCURATE REFERENCE.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF HIGH BLOOD RESULTS. EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100 TO 150 MG/DL. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION DUE TO METER'S RESULTS IS NOT REQUIRED AT THE TIME OF THE CALL ON (B)(6) 2015. BACK TO BACK BLOOD TEST PERFORMED FASTING DURING CALL ON (B)(6) 2015 PRODUCED RESULTS OF 111MG/DL AND 123MG/DL. VERIFIED STORAGE OF PRODUCT IS WITHIN INSTRUCTED SPEC SINCE THEY ARE KEPT IN BEDROOM. TEST STRIP LOT MFR'S EXPIRATION DATE IS 03/31/2018 AND OPEN VIAL DATE IS (B)(6) 2015. RECALL TEST RESULTS PERFORMED FASTING FROM METER MEMORY (DATE / TIME NOT SET): 170MG/DL (B)(6) 2015 07:18AM FASTING YES; 182MG/DL (B)(6) 2015 07:10AM FASTING YES; 182MG/DL (B)(6) 2015 07:10AM FASTING YES; 190MG/DL (B)(6) 2015 07:05AM FASTING YES; 167MG/DL (B)(6) 2015 07:03AM FASTING YES. ADVERSE EVENT NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463765 TRUE RESULT BLOOD GLUCOSE SYSTEM LFR NIPRO DIAGNOSTICS, INC. TRUE RESULT PS2271

Patients

Seq Age Sex Outcome Treatment
1