TRUERESULT
Report
- Report Number
- 1052693-2015-01184
- Event Type
- Malfunction
- Date Received
- July 16, 2015
- Date of Event
- June 22, 2015
- Report Date
- December 11, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USER'S TEST STRIP HAD POOR FILL. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.
PRODUCT NOT YET EVALUATED. (B)(4).
CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CUSTOMER STATES THAT SHE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 121-161MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 01/22/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WAS FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 129MG/DL AND 131MG/DL FASTING. REVIEWED METER MEMORY: SEE SCANNED TABLE.
CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CUSTOMER STATES THAT SHE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 121-161MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 01/22/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WAS FIRST OPENED ON (B)(6) 2015. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 129MG/DL AND 131MG/DL FASTING. REVIEWED METER MEMORY: (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 463751 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PS2204 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |