FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 4924454 · Received July 16, 2015

Report

Report Number
1052693-2015-01184
Event Type
Malfunction
Date Received
July 16, 2015
Date of Event
June 22, 2015
Report Date
December 11, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USER'S TEST STRIP HAD POOR FILL. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.

Additional Manufacturer Narrative · 1

PRODUCT NOT YET EVALUATED. (B)(4).

Description of Event or Problem · 1

CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CUSTOMER STATES THAT SHE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 121-161MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 01/22/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WAS FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 129MG/DL AND 131MG/DL FASTING. REVIEWED METER MEMORY: SEE SCANNED TABLE.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CUSTOMER STATES THAT SHE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 121-161MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 01/22/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WAS FIRST OPENED ON (B)(6) 2015. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 129MG/DL AND 131MG/DL FASTING. REVIEWED METER MEMORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463751 TRUERESULT BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PS2204

Patients

Seq Age Sex Outcome Treatment
1 0 YR