FDA Adverse Event Other Summary report: N

CM LAG SCREW INSERTER - ASIA

MDR report key: 4924354 · Received July 16, 2015

Report

Report Number
9613350-2015-00794
Event Type
Other
Date Received
July 16, 2015
Date of Event
July 3, 2015
Report Date
July 3, 2015
Manufacturer
ZIMMER GMBH
Product Code
LXH
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURER DID NOT RECEIVED DEVICE FOR REVIEW. PICTURES OF THE DEVICES WERE RECEIVED. AS NO LOT NUMBERS WERE PROVIDED FOR THE DEVICES, THE DEVICE HISTORY RECORDS COULD NOT BE REVIEWED. A CAUSE FOR THIS SPECIFIC EVENT CANNOT BE ASCERTAINED FROM THE INFORMATION PROVIDED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE AND AN INVESTIGATION RESULT BE AVAILABLE, THAT CHANGES THIS ASSESSMENT, AN AMENDED MEDICAL DEVICE REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

DEVICES ANALYSIS: AFTER THE STANDARD DECONTAMINATION PROCEDURE PERFORMED AT ZIMMER BIOMET, ONLY THE NAIL AND THE TARGETING GUIDE WAS SHOWING CORROSION TRACES. THE VISUAL EXAMINATION OF THE PARTS REVEALED CORROSION ON AROUND THE SPHERES AND PLUGS IN DIVERSE PLACES ON THE INSTRUMENTS. THE RECEIVED PARTS WERE PHOTOGRAPHED AND THEN CLEANED BY A STANDARD RUST REMOVAL AGENT (15% DECONEX 34 GR). AFTER THE CLEANING WITH THE RUST REMOVAL AGENT THE CORROSION ON THE INSTRUMENT DIMINISHED. THIS WAS DOCUMENTED IN PHOTOGRAPHS. AFTERWARD, A STANDARDIZED AND POPULAR CORROSION TEST WAS PERFORMED ON THE INSTRUMENTS. THE INSTRUMENTS WERE SET FOR 45 MINUTES IN A SALTY SOLUTION. THE SALTY SOLUTION PRODUCES A CORROSIVE ATTACK TO THE SAMPLES (IN THIS CASE THE INSTRUMENT) TO EVALUATE IF THE RUST IS ORIGINATING FROM THE SAMPLE MATERIAL OR NOT. IN THIS CASE, THERE WAS NO APPEARANCE OF NEW CORROSION PRODUCTS (RUST) AFTER THE CORROSION TEST. THIS MEANS THAT THE CORROSION IS NOT ORIGINATING FROM THE INSTRUMENT MATERIAL. REVIEW OF INTERNAL DOCUMENTS: THE INSTRUMENT MATERIAL IS IN ACCORDANCE TO ISO 7153-1, ISO 16061, EN 10088-1, EN 10088-2, EN 10088-3, AND ASTM F899. INSTRUMENT CLEANING IS DESCRIBED IN MANUAL ORTHOPAEDIC SURGICAL INSTRUMENTS. 1. POSSIBLE CAUSES FOR THE REPORTED EVENT ACCORDING TO DFMEA: A) LINE 7: DISCOLORATION/CORROSION FOUND ON THE INSTRUMENTS DUE TO IMPROPER CHOICE OF STERILIZATION METHOD/CYCLE. B) LINE 9: DISCOLORATION/CORROSION FOUND ON THE INSTRUMENTS DUE TO INADEQUATE CLEANING METHOD, INCORRECT CLEANING AGENTS USED. C) LINE 15: DISCOLORATION/CORROSION FOUND ON THE INSTRUMENTS DUE TO CORROSION PRODUCTS PRESENT ON INSTRUMENT. 2. COMPARISON TO INVESTIGATION RESULTS WHETHER IT IS POSSIBLE & JUSTIFICATION: A) POSSIBLE: AFTER THE STANDARD DECONTAMINATION PROCEDURE PERFORMED AT ZIMMER BIOMET, ONLY THE NAIL AND THE TARGETING GUIDE WAS SHOWING CORROSION TRACES. THE VISUAL EXAMINATION OF THE PARTS REVEALED CORROSION ON AROUND THE SPHERES AND PLUGS IN DIVERSE PLACES ON THE INSTRUMENTS. THE RECEIVED PARTS WERE PHOTOGRAPHED AND THEN CLEANED BY A STANDARD RUST REMOVAL AGENT (DECONNEX). AFTER THE CLEANING WITH THE RUST REMOVAL AGENT THE CORROSION ON THE INSTRUMENT DIMINISHED. THIS WAS DOCUMENTED IN PHOTOGRAPHS. AFTERWARD, A CORROSION TEST WAS PERFORMED ON THE INSTRUMENTS. THE INSTRUMENTS WERE SET FOR 45 MINUTES IN A SOLUTION. AFTER THE TEST NO NEW CORROSION WAS FORMED. THIS MEANS THAT THE CORROSION IS NOT ORIGINATING FROM THE INSTRUMENT MATERIAL. B) POSSIBLE: AFTER THE STANDARD DECONTAMINATION PROCEDURE PERFORMED AT ZIMMER BIOMET, ONLY THE NAIL AND THE TARGETING GUIDE WAS SHOWING CORROSION TRACES. THE VISUAL EXAMINATION OF THE PARTS REVEALED CORROSION ON AROUND THE SPHERES AND PLUGS IN DIVERSE PLACES ON THE INSTRUMENTS. THE RECEIVED PARTS WERE PHOTOGRAPHED AND THEN CLEANED BY A STANDARD RUST REMOVAL AGENT (DECONNEX). AFTER THE CLEANING WITH THE RUST REMOVAL AGENT THE CORROSION ON THE INSTRUMENT DIMINISHED. THIS WAS DOCUMENTED IN PHOTOGRAPHS. AFTERWARD, A CORROSION TEST WAS PERFORMED ON THE INSTRUMENTS. THE INSTRUMENTS WERE SET FOR 45 MINUTES IN A SOLUTION. AFTER THE TEST NO NEW CORROSION WAS FORMED. THIS MEANS THAT THE CORROSION IS NOT ORIGINATING FROM THE INSTRUMENT MATERIAL. C) NOT POSSIBLE: THE INVESTIGATION SHOWED THAT THE CORROSION WAS NOT ORIGINATING FROM THE INSTRUMENT MATERIAL ITSELF. BASED ON THE GIVEN INFORMATION AND THE RESULTS OF THE INVESTIGATION, WE COULD IDENTIFY A ROOT CAUSE FOR THIS ISSUE. THE INVESTIGATION SHOWED THAT THE CORROSION WAS NOT ORIGINATING FROM THE INSTRUMENT MATERIAL ITSELF. THE DISCOLORATION ON THE INSTRUMENTS MUST HAVE OCCURRED DUE TO STERILIZED WITH UNDEFINED STERILIZATION CONFIGURATION(E.G.,MIXED ITEMS) OR INADEQUATE CLEANING METHOD (INCORRECT CLEANING AGENTS USED). IT IS POSSIBLE THAT THE RUST HAS BEEN DEPOSITED FROM OTHER INSTRUMENTS. THE NEED FOR CORRECTIVE MEASURES IS NOT INDICATED AND ZIMMER (B)(4) CONSIDERS THIS CASE AS CLOSED. ZIMMER'S REFERENCE NUMBER OF THIS FILE IS (B)(4).

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED ON 07/27/2015. WHERE LOT NUMBERS WERE RECEIVED FOR THE DEVICE, THE DEVICE HISTORY RECORDS WERE REVIEWED AND FOUND TO BE CONFORMING. AS SOON AS ADDITIONAL INFORMATION BECOMES AVAILABLE AND/OR AN INVESTIGATION RESULT BE AVAILABLE, AN AMENDED MEDICAL DEVICE REPORT WILL BE SUBMITTED. ZIMMER'S REFERENCE NUMBER OF THIS FILE IS (B)(4).

Additional Manufacturer Narrative · 1

DHR REVIEW: DHR RECORDS WERE REVIEWED AND FOUND TO BE CONFORMING. TREND ANALYSIS: NO TREND IDENTIFIED. COMPATIBILITY CHECK: THE COMPATIBILITY CHECK COULD NOT BE PERFORMED AS ONLY ONE PRODUCT WAS REPORTED TO US. THE PRODUCT COMPATIBILITY CHECK IS NOT RELEVANT FOR ONE PRODUCT ONLY. THE AFFECTED PRODUCTS WERE NOT USED IN THE SAME SURGERY. REVIEW OF INCOMING INFORMATION: IT IS REPORTED THAT THE INSTRUMENTS PRESENTS MASSIVE RUST AFTER CLEANING. 5 PICTURES WERE AVAILABLE FOR INVESTIGATION, SHOWING STAINING OR "INCRUSTATION" ON THE INSTRUMENTS. NO OTHER DOCUMENTS WERE PROVIDED. REVIEW OF INTERNAL DOCUMENTS: INSPECTION PLAN: VISUAL INSPECTION PERFORMED AQL 0.65%. THE INSTRUMENT MATERIAL IS IN ACCORDANCE TO STANDARDS. INSTRUMENT CLEANING IS DESCRIBED IN MANUAL ORTHOPAEDIC SURGICAL INSTRUMENTS. BASED ON THE GIVEN INFORMATION AND THE RESULTS OF THE INVESTIGATION, WE COULD IDENTIFY A ROOT CAUSE FOR THIS ISSUE. DEDUCTED FROM THE EVENT DESCRIPTION AND AFTER PICTURES EXAMINATION THE FOUND ISSUE IS CAUSED BY STAINING. STAINING IS A SURFACE DEPOSIT ON INSTRUMENTS THAT IS MOST OFTEN MISTAKEN FOR RUST. STAINING AND WATER SPOTS WERE FOUND ON THE INSTRUMENTS DUE TO IMPROPER CLEANING CYCLE. WATER SPOTS CAN BE CAUSED BY A TOO HIGH CONCENTRATION OF CLEANING AGENTS OR MINERALS OR ORGANIC RESIDUALS. THE USE OF CITY WATER DURING THE WASHING AND STERILIZATION PROCESS ALSO PRODUCES FLECKS. GENERALLY, PURIFIED WATER SHOULD BE USED IN THE LAST TWO RINSE STEPS BY WASHING AND IN THE AUTOCLAVE PROCESS. ALSO OVERLOADED INSTRUMENT TRAYS IN THE AUTOCLAVE OR INSTRUMENT CONTACT SURFACES CAN CAUSE WATER SPOTS. ORANGE/BROWN STAIN: THE PROBLEM IS MOST OFTEN A PHOSPHATE LAYER (BROWN TO LIGHT ORANGE) ON THE INSTRUMENT, WHICH DEVELOPS AS A RESULT OF ANY OF THE FOLLOWING CAUSES: WATER SOURCES, DETERGENTS USED TO WASH AND CLEAN INSTRUMENTS, SURGICAL WRAPPINGS, COLD STERILIZATION SOLUTIONS, OR DRIED BLOOD. THE NEED FOR CORRECTIVE MEASURES IS NOT INDICATED AND ZIMMER (B)(4) CONSIDERS THIS CASE AS CLOSED. ZIMMER'S REFERENCE NUMBER OF THIS FILE IS (B)(4).

Description of Event or Problem · 1

IT WAS NOW REPORTED THAT MASSIVE CORROSION WAS DETECTED ON CM LAG SCREW INSERTER - (B)(4) DURING CLEANING.

Description of Event or Problem · 1

IT WAS REPORTED, THAT MASSIVE CORROSION WAS DETECTED ON AN UNKNOWN ZNN INSTRUMENT DURING CLEANING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463777 CM LAG SCREW INSERTER - ASIA ZIMMER NATURAL NAIL SYSTEM CEPHALOMEDULAR LXH ZIMMER GMBH NA 14015900

Patients

Seq Age Sex Outcome Treatment
1 Other