FDA Adverse Event Malfunction Summary report: N

INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT

MDR report key: 4924104 · Received July 19, 2015

Report

Report Number
9611451-2015-00318
Event Type
Malfunction
Date Received
July 19, 2015
Report Date
June 22, 2015
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BZE
PMA / PMN Number
K103767
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CB
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6). DEVICE MFR DATE: (B)(6) (B)(6) 2014 (B)(6) (B)(6) 2015 (B)(6) (B)(6) 2015 METHOD: EIGHT OF THE NINE COMPLAINT (B)(4) INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUITS WERE RETURNED TO FPH IN (B)(6) AND WERE PRESSURE TESTED FOR LEAKS. THE UNIT WITHOUT LOT NUMBER, AS SPECIFIED ABOVE, WAS NOT RETURNED TO FPH FOR INSPECTION. AN ATTEMPT WAS ALSO MADE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE REPORTED COMPLAINT BUT NO RESPONSE WAS RECEIVED FROM THE HEALTHCARE FACILITY. RESULTS: NO FAULT WAS FOUND WITH THE RETURNED (B)(4) INFANT BREATHING CIRCUITS DURING PRESSURE TEST. THE TEST RESULTS WERE WITHIN SPECIFICATION. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBERS 141022, 150113, 150129 AND 150211. CONCLUSION: THE LEAK REPORTED BY THE HEALTHCARE FACILITY WAS NOT REPLICATED DURING INSPECTION. WE ARE THEREFORE UNABLE TO DETERMINE WHAT MAY HAVE CAUSED THE REPORTED PROBLEM. ALL RT266 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE AND FLOW TESTED PRIOR TO DISTRIBUTION, AND THOSE THAT FAIL ARE REJECTED. THE SUBJECT RT266 INFANT BREATHING CIRCUIT WOULD HAVE MET THE REQUIRED SPECIFICATIONS AT THE TIME OF PRODUCTION. THIS SUGGESTS THE LEAK OCCURRED POST PRODUCTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT266 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT STATE THE FOLLOWING: - "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." - "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." - "SET APPROPRIATE VENTILATOR ALARMS."

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT NINE RT266 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUITS WERE FOUND LEAKING AROUND THE PRESSURE PORT DUE TO "IMPROPER ATTACHMENT". THIS WAS OBSERVED DURING USE ON PATIENTS. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467946 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LIMITED RT266 SEE H10 MANUFACTURER NARRATIVE

Patients

Seq Age Sex Outcome Treatment
1