EVIS EXERA III BRONCHOVIDEOSCOPE
Report
- Report Number
- 2951238-2015-00300
- Event Type
- Death
- Date Received
- July 17, 2015
- Date of Event
- June 10, 2015
- Report Date
- June 18, 2015
- Manufacturer
- OLYMPUS MEDICAL SYSTEM CORPORATION
- Product Code
- EOQ
- PMA / PMN Number
- K112846
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- RISK MANAGER
Narratives
THE DEVICE REFERENCED IN THIS REPORT HAS NOT YET BEEN RETURNED TO OLYMPUS FOR EVALUATION. A REVIEW OF THE INSTRUMENT HISTORY SHOWS THAT THE DEVICE WAS PURCHASED ON 10/25/2013 AND WAS LAST SERVICED ON 02/04/2015. THE EXACT CAUSE OF THE PATIENT'S OUTCOME COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE AT A LATER TIME, THIS REPORT WILL BE SUPPLEMENTED.
OLYMPUS RECEIVED A MEDWATCH REPORT #(B)(4) STATING "BRONCHOSCOPE IN USE WHEN AN AIRWAY FIRE OCCURRED DURING THE USE OF A YAG LASER TO ABLATE AN UPPER LOBE ENDO-BRONCHIAL TUMOR." OLYMPUS FOLLOWED UP WITH THE USER FACILITY TO OBTAIN ADDITIONAL INFORMATION REGARDING THE REPORTED EVENT AND WAS INFORMED THAT THE EVENT OCCURRED DURING A LASER BRONCHOSCOPY WITH ABLATION OF A LEFT LUNG TUMOR PROCEDURE. A FIRE OCCURRED IN THE THROAT/ AIRWAY OF THE PATIENT WHILE USING THE DEVICE IN CONNECTION WITH AN AMS LASER AND AN AESTIVA ANESTHESIA MACHINE, THE FIRE MELTED THE TIP OF THE SCOPE ABOUT 5 INCHES FROM THE DISTAL TIP. THE PATIENT WAS HOSPITALIZED POST SURGERY FOR ONE DAY AND WAS THEN TRANSFERRED TO THE BURN UNIT. PATIENT EXPIRED AT THE BURN UNIT ON (B)(6) 2015. IT IS UNKNOWN AT THIS TIME WHAT LED TO THE PATIENT'S OUTCOME. NO ADDITIONAL INFORMATION WAS PROVIDED AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 466183 | EVIS EXERA III BRONCHOVIDEOSCOPE | EVIS EXERA III BRONCHOVIDEOSCOPE | EOQ | OLYMPUS MEDICAL SYSTEM CORPORATION | BF-1TH190 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | Death | AESTIVA ANEATHESIA MACHINE MODEL AESTIVA| CLV-190 EVIS EXERA III VIDEO XENON LIGHT SOURCE| CV-190 EVIS EXERA III VIDEO SYSTEM| KTP/YAG LASER: LASERSCOPE MODEL 804| AESTIVA ANEATHESIA MACHINE MODEL AESTIVA| CLV-190 EVIS EXERA III VIDEO XENON LIGHT SOURCE| KTP/YAG LASER: LASERSCOPE MODEL 804| CV-190 EVIS EXERA III VIDEO SYSTEM |