FDA Adverse Event Death Summary report: N

EVIS EXERA III BRONCHOVIDEOSCOPE

MDR report key: 4922961 · Received July 17, 2015

Report

Report Number
2951238-2015-00300
Event Type
Death
Date Received
July 17, 2015
Date of Event
June 10, 2015
Report Date
June 18, 2015
Manufacturer
OLYMPUS MEDICAL SYSTEM CORPORATION
Product Code
EOQ
PMA / PMN Number
K112846
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE REFERENCED IN THIS REPORT HAS NOT YET BEEN RETURNED TO OLYMPUS FOR EVALUATION. A REVIEW OF THE INSTRUMENT HISTORY SHOWS THAT THE DEVICE WAS PURCHASED ON 10/25/2013 AND WAS LAST SERVICED ON 02/04/2015. THE EXACT CAUSE OF THE PATIENT'S OUTCOME COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE AT A LATER TIME, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

OLYMPUS RECEIVED A MEDWATCH REPORT #(B)(4) STATING "BRONCHOSCOPE IN USE WHEN AN AIRWAY FIRE OCCURRED DURING THE USE OF A YAG LASER TO ABLATE AN UPPER LOBE ENDO-BRONCHIAL TUMOR." OLYMPUS FOLLOWED UP WITH THE USER FACILITY TO OBTAIN ADDITIONAL INFORMATION REGARDING THE REPORTED EVENT AND WAS INFORMED THAT THE EVENT OCCURRED DURING A LASER BRONCHOSCOPY WITH ABLATION OF A LEFT LUNG TUMOR PROCEDURE. A FIRE OCCURRED IN THE THROAT/ AIRWAY OF THE PATIENT WHILE USING THE DEVICE IN CONNECTION WITH AN AMS LASER AND AN AESTIVA ANESTHESIA MACHINE, THE FIRE MELTED THE TIP OF THE SCOPE ABOUT 5 INCHES FROM THE DISTAL TIP. THE PATIENT WAS HOSPITALIZED POST SURGERY FOR ONE DAY AND WAS THEN TRANSFERRED TO THE BURN UNIT. PATIENT EXPIRED AT THE BURN UNIT ON (B)(6) 2015. IT IS UNKNOWN AT THIS TIME WHAT LED TO THE PATIENT'S OUTCOME. NO ADDITIONAL INFORMATION WAS PROVIDED AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
466183 EVIS EXERA III BRONCHOVIDEOSCOPE EVIS EXERA III BRONCHOVIDEOSCOPE EOQ OLYMPUS MEDICAL SYSTEM CORPORATION BF-1TH190 N/A

Patients

Seq Age Sex Outcome Treatment
1 87 YR Death AESTIVA ANEATHESIA MACHINE MODEL AESTIVA| CLV-190 EVIS EXERA III VIDEO XENON LIGHT SOURCE| CV-190 EVIS EXERA III VIDEO SYSTEM| KTP/YAG LASER: LASERSCOPE MODEL 804| AESTIVA ANEATHESIA MACHINE MODEL AESTIVA| CLV-190 EVIS EXERA III VIDEO XENON LIGHT SOURCE| KTP/YAG LASER: LASERSCOPE MODEL 804| CV-190 EVIS EXERA III VIDEO SYSTEM