FDA Adverse Event Malfunction Summary report: N

MCK TIBIAL ONLAY INSERT-SZ 5-8MM

MDR report key: 4922954 · Received July 17, 2015

Report

Report Number
3005985723-2015-00127
Event Type
Malfunction
Date Received
July 17, 2015
Date of Event
June 22, 2015
Report Date
June 22, 2015
Manufacturer
MAKO SURGICAL CORP.
Product Code
NPJ
PMA / PMN Number
K090763
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN EVENT REGARDING DAMAGED INSERT DUE TO SEATING ISSUE INVOLVING A MAKO INSERT WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD & RESULTS: DEVICE EVALUATION: VISUAL INSPECTION CONFIRMED THE INSERT WAS DAMAGED DUE TO IMPLANTATION. MEDICAL RECORDS RECEIVED AND EVALUATION: NO MEDICAL RECORDS WERE RECEIVED FOR REVIEW WITH A CLINICAL CONSULTANT. DEVICE HISTORY REVIEW INDICATED THE DEVICES ACCEPTED INTO FINAL STOCK FROM THE REPORTED LOT WERE FREE FROM DISCREPANCIES. COMPLAINT HISTORY REVIEW: THERE HAVE BEEN 2 OTHER EVENTS FOR THE LOT REFERENCED. CONCLUSION: THE INVESTIGATION CONFIRMED THAT THE DEVICE WAS DAMAGED DURING ATTEMPT AT IMPLANTATION. THERE IS NO INDICATION AT THIS TIME THAT THE DESIGN, MATERIALS, OR MANUFACTURING OF THE SUBJECT DEVICE CONTRIBUTED TO THE EVENT. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS INVESTIGATION WILL BE REOPENED.

Additional Manufacturer Narrative · 1

WHEN COMPLETED, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

TIBIAL INSERT WAS DAMAGED WHILE SURGEON WAS TRYING TO SEAT IT. A SECOND INSERT WAS OPENED AND IMPLANTED. CASE WAS SUCCESSFUL AND NO UNUSUAL CIRCUMSTANCES WERE EXPERIENCED.

Description of Event or Problem · 1

TIBIAL INSERT WAS DAMAGED WHILE SURGEON WAS TRYING TO SEAT IT. A SECOND INSERT WAS OPENED AND IMPLANTED. CASE WAS SUCCESSFUL AND NO UNUSUAL CIRCUMSTANCES WERE EXPERIENCED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467258 MCK TIBIAL ONLAY INSERT-SZ 5-8MM IMPLANT NPJ MAKO SURGICAL CORP. 12340215

Patients

Seq Age Sex Outcome Treatment
1 Other